Page 2
Limerick, Inc., is proud that PJ's Comfort breast pumps are designed by breastfeeding mothers for breast-feeding mothers. Our company is owned by a mother daughter team who are certified lactation consultants with over 20 years of experience assisting mothers who use our pumps in the workplace.
Page 3
7. Cleaning accessory kit ......See page 27 *This overview is not intended to provide complete user instructions. Read the entire manual before operating pump. The Owner’s Manual can also befound on the Limerick Inc. website www.limerickinc.com Owner’s Manual revision: October 2019...
Page 4
Symbols and their meaning If the accessory kit is not labeled sterile see page 10 for instructions on sanitizing the accessory kit. Sanitizing the accessory kit needs to be done before the first time you use it. After the first time you use the pump, the accessory kit will only need a simple cleaning procedure found on page 27.
10. Make sure the voltage of the pump is compatible with the pump. If you plan to use this pump in another country where voltage and outlet plugs may differ, a plug adapter may be required. Please contact Limerick, Inc., for recommendations.
Page 7
8. If the filter becomes plugged, replace the filter. Do not tamper with the filter. Please refer to page 23 “Caution” to prevent clogging. 9. DO NOT ATTEMPT TO REPAIR PJ's Comfort hospital grade electric breast pump. There are no user-serviceable parts inside.
Introduction to PJ’s Comfort Breast Pumps Pump Description The PJ’s Comfort hospital grade multi-user breast pump is intended to express and collect milk from a lactating women’s breast. This action helps to alleviate engorgement of the breast, maintain the woman’s ability to lactate, and provide a mother’s milk for future feedings when separation of the mother and baby occur.
Page 11
Benefits of using PJ’s Comfort electric breast pump • Establish and maintain milk production by emptying the breast if breast-feeding is temporarily interrupted • Increase milk production • Continue to provide infant with breast milk when mother and infant are separated Page 31 has a list of optional accessories and replacements parts.
Accessory Kit Sanitizing The filter, tubing, connector, clamp and “Y” pieces do not need to be sanitized as air is all that flows through these parts. The filter is designed to last the duration of your current breast feeding goals Please refer to page 23 “Caution” to prevent clogging. Page 10...
Controls and Assembly Pump Assembly 1. Place pump on a level surface. 2. Make sure the power/cycle knob (1) is turned off - all the way to the left. 3. Turn vacuum adjustment knob ( 2 ) all the way to the left. 4 Insert the two storage container holders (6) into the storage container holder slots ( 7 ) on either side of the base of the pump.
Page 15
Purpose of Pump Controls and Features 5 pages) The pump has two knobs which control the operation. Rotating the knobs adjust the settings. When a knob is adjusted, the time display automatically changes to show the setting value, then switches back to time shortly after the adjustment has stopped.
Page 16
Purpose of Pump Controls and Features con’t) Since each mother/baby couple’s breast-feeding experience is different, the adjustment of these knobs will be an individual choice. The goal is to adjust the suction rhythm (cycle) to what best mimics your baby’s sucking pattern and to adjust the suction strength (vacuum) to a comfortable setting.
Page 18
This fault means that the vacuum increased to a level higher than expected. This can only happen if there is an internal malfunction. If resetting does not restore normal operation contact Limerick as the pump may need to be returned for repair. Flashing 10, 20 Lights: This error code means an internal malfunction has occurred, contact Limerick as the pump may need to be returned for repair.
Page 19
Purpose of Pump Controls and Features con’t) Other situations that may occur without a fault message Situation “A” If tubing becomes kinked or milk back flows into the filter the pump will lose vacuum and pulsate quickly. You will either need to un-kink tubing or change the filter if milk is in the filter as the filter can not be cleaned.
Page 25
Your PJ’s Comfort breast pumps are equipped with storage containers for collection and storage of milk. Also supplied are caps to hold the white seal for storage of breast milk (parts 20-21 on page 7). See page 31 for information on purchasing additional storage containers and accessories.
Page 26
Ways to Use PJ’s Comfort Breast Pumps to Express Milk Page 24...
Pump and Accessory Care Pump Care and Cleaning When pump is shared by multiple users, such as in the hospital, wipe unit with a damp cloth containing a germicidal detergent. Dry with a soft cloth. Wash holders in detergent, using germicide when more than one person uses them.
Accessory Kit Cleaning Instructions Before the first time you use your pump, follow the sanitizing instructions on pages 10-11. After each use: 1. Remove tube connector/clear tubing assembly 10-13 from both breast cup storage container assemblies (14- 16). DO NOT WASH FILTER (9), TUBING (10 &...
14 With the pump still running, place both of the breast cups onto a hard smooth surface, such as a table top. 15 Both cups should be compressing and releasing. 16 If pump is still not working call Limerick Customer Service. Page 28...
Page 31
You may also try calling the location where you obtained your breast feeding prod- uct. Before use visually inspect pump for cracks, chips or deterioration. In the event that damage too the device occurred, discontinue use and return to Limerick. Limerick will perform all preventive maintenance and recalibration when necessary. No serviceable parts.
Page 32
Transportation Information Temperature between 40 F degree to 130 F degrees Humidity: 75% Altitude 35,000 feet Page 30...
Ordering Replacement Parts or Optional Accessories Replacement parts can be ordered from Limerick For a catalog, call 877-546-3742 or visit www.limerickinc.com Refer to the diagram on pages 6-7 to identify parts. Filter with tubing Breast cup assembly Cleaning brush Storage container...
(Kit contents and power supply not included). Warranty is non-transferable This product is warranted by Limerick Inc. to the original purchaser to be free from defects in material and workmanship. In the event of a defect, Limerick will repair, or at Limerick’s option, replace this product, without charge for such replacement...
Need help?
Do you have a question about the PJ's Comfort and is the answer not in the manual?
Questions and answers