4 Calibration Check
The calibration of the Timbermaster
is checked by holding the electrode
needles across the exposed wires of
the 'calcheck' device (supplied) or
across the terminals of the Protimeter
Checkbox (optional accessory).
When checking the calibration, the A
scale should be selected and the tem-
perature probe must be disconnected.
Correctly calibrated Timbermaster
will register a (%MC) value in the
range of 17.8 to 18.3.
5 Care and Maintenance
When the Timbermaster is not in use,
keep it in its pouch together with its
accessories. Store the kit in a stable,
dust-free environment out of direct
sunlight. Remove the batteries from
the instrument if it is to be stored for
periods of more than four weeks, or
when the low battery power symbol
appears on the display. Check the
condition of accessories used with the
Timbermaster instrument on a regular
basis and replace them if they become
worn or damaged.
The information contained in this leaflet
is given in good faith. As the method of
use of the instrument (and its accesso-
ries) and the interpretation of the read-
ings are beyond the control of the
manufacturers, they cannot accept
responsibility for any loss, consequen-
tial or otherwise, resulting from its use.
USA
1100 Technology Park Drive
Billerica, MA 01821-4111
www.gesensing.com/protimeterproducts
Ireland
Shannon Industrial Estate
Shannon, County Clare
www.gesensing.com/protimeterproducts
®
Protimeter Timbermaster
is a registered trademark of GE
INS5601, Rev. A
January 2006
User Instructions
4
Need help?
Do you have a question about the Protimeter Timbermaster and is the answer not in the manual?
Questions and answers