7. EXAMINATION

7.1. IMPORTANT INFORMATION

Do not use the device while it is being charged.
REMEMBER
Do not perform an examination in a room where the
temperature is lower than 20˚C or higher than 25˚C�
Due to changes in breast physiology during the men-
strual cycle, if you wish to compare the subsequent
results of the same patient, select the examinations
performed on a similar day of the menstrual cycle�
Prior to the examination, please make sure that the patient:
1. On the day before the examination:
avoided procedures which may have increased blood circu-
l
lation in the breast area (physiotherapy, massages, etc�)�
2. On the day of the examination:
avoided procedures which may have increased blood circu-
l
lation in the breast area (physiotherapy, massages, etc�);
did not take a bath or a shower within approximately two
l
hours before the examination;
did not use peelings on her breasts and/or neckline and did not
l
perform armpit depilation, as this may have irritated her skin;
refrained from sources of heat or cold, such as heaters or air
l
conditioners, for at least thirty minutes before and during the
examination;
avoided intense physical effort for approximately two hours
l
prior to the examination;
did not use cosmetics and did not wet the skin on her
l
breasts within two hours prior to the examination – her skin
should be clean and dry;
did not have a sauna or a steam bath�
l