Copyright This document is the exclusive property of medin Medical Innovations GmbH. No part of this document may be copied or otherwise reproduced or stored in any form on electronic media without prior written consent of the company, except in the case of express authorisation from the company. Contact the manufacturer for further information.
Page 3
Each guarantee which cannot be fully excluded is limited to the duration of this warranty, if permitted by applicable law. The service life of the device is defined as 8 years. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Power Supply 2.6. Atmosphere and Environment 2.7. Dimensions and Weight 2.8. Overview of the General Technical Description 2.9. Accessories for medin medinSINDI ® REF 1080 3. Unpacking and Set-Up 3.1. Mounting and Setting Up the Parts 3.2. Connection to Gas Supply 3.3.
Please make a note of the serial number on the identification plate on the back of the device for your records. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com...
Important Note - Haemodynamic Monitoring The device may only be used with simultaneous haemodynamic monitoring of the patient. Only trained hospital staff are authorised to use the device on patients. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
When connected to a central gas supply system: use pressure reducers at the gas supply tubes to compensate for pressure differences (see list of accessories on page 29) +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com...
Gas connections: for DISS supply (alternative to NIST) Oxygen sensor for FiO measurement - sensor type MLF16 (unleaded) Overpressure valve 80 mbar +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Vibrations: not in locations exposed to constant vibration +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
® patient. Only the patient tubes, the Medijet and the masks and prongs used are used in the direct vicinity of the patient. ® 2.9. Accessories for medin medinSINDI REF 1080 Description Optional NIST connection 2020/2030 Air and oxygen supply tubes DISS...
Attach air gas supply tube (b/w) to central gas supply or full gas cylinder (3.5 – 6 bar) Attach oxygen gas supply tube (white) to central gas supply or full gas cylinder (3.5 – 6 bar) +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com...
Disconnect the power supply unit from the mains during thunderstorms or if the device is not used for a longer period of time. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Page 13
Do not insert the power supply plug of the medinSINDI ® into ceiling outlets since, due to its weight, it can be pulled out of the outlet in this case. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
After the device start, activate "alarm silent". Then "CPAP alarm off" appears in the upper right- hand part of the display. Now the therapy can be started and the flow as well as the concentration can be adjusted accordingly. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Soft-press buttons for alarm adjustment (see chapter 4.3.4 and chapter 5) Monitor with integrated 96 x 61 mm LCD display, resolution 240 x 128 dots b/w, including illumination Display for amount of driving flow +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Training after installation by authorized sales and service partners! For use by trained professionals only! After long idle times, oxygen calibration must be performed - see service menu oxygen calibration (page 20). +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Starting with software version 1.022 the current CPAP low and CPAP high values change in conjunction. Starting with software version 1.0124 Alarms are prioritized with melody +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
FiO values are deleted. Starting with software version 1.023, trending is automatically discontinued in the event of an alarm. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
CPAP buttons must be held down for another 3 seconds. The CPU will then start the “SYSTEM TEST” and open the service menu after this test. The display then appears. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com...
Figure 9: Oxygen calibration Important Note – Oxygen Calibration Wait until the ADC VALUE is STABLE! Then confirm. After the calibration is completed, the service menu reappears. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
To be able to use it, it must be activated in the particular device. In order to do this, the service menu must be opened and then the item "Acoustic Alarm off (not) possible" must be selected. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
O cell. The CPU works with "Flash Memory". At the customer's request, medin can update the programme with the most recent software version at any time in the Puchheim plant. To close this menu, hold down the “Confirm FiO ”...
Monitoring of the oxygen concentration set. Activation with "CONFIRM FiO2" necessary Battery In the event of a shortfall of the minimum charge, there will be a 1-second beep each minute +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
6.5. Power Supply Unit (Cleaning, Maintenance, Repairs and Modifications) Cleaning: Prior to cleaning, the power supply unit of the medinSINDI ® must be disconnected from the mains. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Note – Device Passport For maintenance and after each opening, the device passport must be filled out and a new safety seal must be affixed. (see appendix) +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Only for sindi Flow outlet Flow outlet Oxygen Oxygen Pneumatic diagram Pneumatic diagram Medin, Medical Innovations München Medin, Medical Innovations München Medin Blender 1085 Medin Blender 1085 Gert Kniewasser Gert Kniewasser file: m/entwicklung/1085pneumatic > May 2006 file: m/entwicklung/1085pneumatic > May 2006 Figure 14: Construction of the monitor and mixer and pneumatic diagram of the mixer ®...
Order number (3,11)EN temperature for labelling the EN 980:2008 980_2003 range of the product (5.7.3) device identification +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017 28 von 36 D-82140 Olching www.medingmbh.com...
IEC 61000-4-8 Note: is the a.c. supply voltage prior to application of the test level +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017 31 von 36 D-82140 Olching www.medingmbh.com...
Page 32
® necessary, such as re-orienting or relocating the medinSINDI Over the frequency range 150 kHz - 80 MHz field strengths should be less than 3 V/m. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
Page 33
At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations. Electromagnetic propagation is Note 2: affected by absorption and reflection from structures, objects and people. +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017...
/ (dis)able Oxygen high low / ok / alarm Gas supply pressure ok / Air/O +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017 34 von 36 D-82140 Olching www.medingmbh.com...
Page 35
- 2015 electromagnetic compatibility - 2016 formatting 2017 new layout, update pneumatic properties +49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017 35 von 36 D-82140 Olching...
Page 36
+49 8142 448 460 medin Medical Innovations GmbH OP_medinSINDI EN Adam-Geisler-Str. 1 info@medingmbh.com Rev 14 Stand 30.10.2017 36 von 36 D-82140 Olching www.medingmbh.com...