5.7.1.1.1 International Calibration Factor (ICF)
Option
The International
Calibration Factor (ICF)
is a factor adjusted to all
absorbance readings.
When the ICF is
changed, all stored
curves are deleted.
Change of International
Calibration Factor is
disabled when there are
tests in the test list.
5.7.1.2 Sample Database Settings
Option
The various options
allow the user to
customize the field
names used in the
sample database. Click
on the checkbox to
uncheck the selection;
only selections that are
checked boxes will be
applied. Click the OK
button once all the
selections have been
made.
NOTE: The Sample ID
field and the Keyword
field must be used.
ChroMate
The allowed range for the ICF is 0.9 to 1.1 (10% change up
or down). If the ICF is out of the allowable range, an error
message will display.
If the ICF is different than 1, a string is added to the note
field of the result record showing the factor's value
(ICF=1.030).
Calibration tab shows result and ICF=X.XXX in the Note
field of each record (when ICF != 0).
The ICF is also printed on calibration report for each record
(when ICF != 0 and the Note field is checked for the
"Report Output").