Prostat PGA-710B Getting Started Operations Manual

Autoanalysis system
Hide thumbs Also See for PGA-710B:

Advertisement

autoanalysis system
pGa-710B
Getting Started Operations Guide

Advertisement

loading
Need help?

Need help?

Do you have a question about the PGA-710B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Prostat PGA-710B

  • Page 1 Getting Started Operations Guide...
  • Page 3: Table Of Contents

    Generating a Detailed Report Copyright © 2013 by Prostat® Corporation. All rights reserved. Printed in the United States of America. No part of this manual may be used or reproduced in any manner whatsoever without written permission. For information contact Prostat Corporation, 1072 Tower Lane, Bensenville, IL 60106 USA Prostat is the registered trademark of Prostat®...
  • Page 4: System Requirements

    • Microsoft® DirectX 9b • Instrument input limits to ± 2 volts. note: Providing greater than ± 2 volts to the PGA-710B Autoanalyzer will void the warranty. For appropri- ate adapters or cables, please contact Prostat Corporation or your Prostat Authorized Reseller.
  • Page 5 PGA-710B Autoanalysis System b. installing the Autoanalysis Application Software nOTE: DO NO connect your PGA-710B Device to your computer or laptop until installation is complete. nOTE .NET Framework 4 is required to run the Autoanalysis Application Software. If .NET Framework 4 is not currently installed, the Autoanalysis Application Software Installation wizard will prompt you to install from the disk.
  • Page 6: Connecting The Pga-710B Autoanalysis System To Your Computer

    C. Connecting the PgA-710b Autoanalysis System to your computer Before you start using or connect the PGA-710B, make sure to charge its battery for at least 8 to 14 hours using the supplied AC/DC converter. Estimated operating battery life is approximately 8 to 10 hours of continuous operation without connection to USB port, or supplied battery charger.
  • Page 7: Adjust Or Confirm The "Zero" Reference For The Pga-710B System

    PGA-710B Autoanalysis System At this point, the “Save and Erase” LED will be On indicating the PGA-710B Autoanalyzer’s battery is re- ceiving a charging voltage from the computer’s USB. nOTE: This charge is not sufficient to completely charge the battery during normal op- erations.
  • Page 8: Assembling Your Autoanalysis System To The Field Meter And Charge Plate Monitor

    Monitor. Attach the Field Meter to a tested earth ground. 2. From the Edit Menu, click on Device Con- trol. Do the following: 2. Connect the PFM-711A to the PGA-710B using the provided analog cable. a. Click Set Zero b. Click Set Device Time c.
  • Page 9: Capture And Analyze Data

    PGA-710B Autoanalysis System APPliCATiOn The PGA-710B records and converts analog input from a field meter instrument to digital data, then transmits that data to your computer via USB, or stores it for future use. Anything your field meter measures will be transmitted via the analog cable to the PGA- 710B, then to the computer where the dynamic measurements are displayed on the screen.
  • Page 10 PGA-710B Autoanalysis System Step 11 Step 12 Voltage generation Add Minimum or Maximum Data Peaks 1. Trimmed portions of the chart are eliminat- 1. Click the Adjust button. This allows addition ed from the analysis or removal of minimum or maximum peaks from the analysis.
  • Page 11: Generating A Detailed Report

    PGA-710B Autoanalysis System G. generating a Detailed Report TiP: There should be an equal number of alternating minimum and maximum peaks. ±3 Sigma in Step 12 must be completed before generating a report (Step 13). Step 13 Step 14 generating a Report Entering information on Report 1.