Vertical Mount - Siemens Milltronics HydroRanger 200 Instruction Manual

Heated enclosure
Table of Contents

Advertisement

Wiring
Warning: Hazardous voltage present on terminal during
operation.
Notes:
• Recommended torque on terminal clamping screws: 0.56 – 0.79 Nm (5–7 in/lbs)
• The equipment must be protected by a 15 A fuse, or a circuit breaker in the
building installation.
• A circuit breaker or switch in the building installation, marked as the disconnect
switch, must be in close proximity to the equipment and within easy reach of
the operator.
• Do not connect level monitor to enclosure terminals.
Important: All safety and operating information in this manual apply to
the Heated Enclosure only. Consult instrument manual for
safety and operating instructions.

Vertical Mount

strip heater
115Vac, 120VA
L1
L2/N
Note: do not wire
level monitor to
Heated Enclosure
terminal.
7ML19981FY01
L 1
L 2 /N
terminal block
ground lug terminal
Heated Enclosure – INSTRUCTION MANUAL
thermostat
level monitor cable entry:
refer to individual product
for power consumption.
COVER NOT SHOWN
Milltronics
Level
Monitor
enclosure
Page 7