Thank you for purchasing the CareMaster. This product is designed to clean and lubricate the
handpiece easily and thouroughly. Please read this Instructions For Use carefully before use and
file for future reference.

Safety Information

Read these safety precautions before use and operate the product properly.
The indicators are to allow you to use the product safely and prevent danger and harm to you
and others. These are classified by degree of danger, damage and seriousness. All indicators
concern safety, and be sure to follow them
WARNING!
(RISk OF InjURy)
(TO PReVenT DAMAge)
Do not install, use or fill this product with Maintenance Oil in a room with a risk of explosion or near open flames.
Do not install or use the product in direct sunlight or where the temperature exceeds 40 °C (104 °F).
Do not use this product for purposes other than the maintenance (cleaning and lubrication) of instruments
described in this Operating Manual.
Provide adequate ventilation. If odours cause your concern, provide additional ventilation immediately.
Do not handle the Power Cord with wet hands as this may result in an electric shock.
Be careful not to spill water onto the Control Unit as this may result in a fire or an electric shock due to a short-
circuit.
Do not disassemble or alter the Control Unit. For service or repair contact your authorised dealer.
If the Control Unit smokes or smells of burning, turn off the power immediate and disconnect the power plug.
Contact your dealer.
Do not drop the Control Unit. Place the Control Unit on a flat surface.
Use a Fuse of specified rating. (120V: T800mAL250V, 220V/240V: T400mAL250V)
Always autoclave handpieces after cleaning and lubricating with this product.
The system functions normally in an environment where the temperature is at 10- 40°C ( 50-104°F ), humidity
at 25-70% RH, atmospheric pressure at 500-1060hPa and no moisture condensation in the Control Unit. Use at
outside these limits may cause malfunction.
Store the system in a place where the temperature is at -10-50°C ( 14-122°F ), humidity at 10-85% RH,
atmospheric pressure at 500-1060 hPa and the Unit is not subject to air with dust, sulfur or salinity.
When installing this product, provide space of approximately 10 cm around the Control Unit.
Place a collector such as a tray under the Control Unit.
The air supply to this device (Air Compressor etc.) and Air Filter need to be clean and free from moisture. Drain
the Air Compressor and Air Filter at least once a week. Moisture from the tank could mix with the Maintenance Oil
and defeat the purpose of this device.
Remove the Air Tube and the Power Cord if the Control Unit is not to be used for a long time.
ATTENTION!
explanations,
without risk to
persons or objects
DANGER
WARNING
CAUTION
General
Thermal washer
disinfectable
3
135°C
Sterilizable
up to the stated
temperature