90
Diatron MI PLC
10.2.7.Serial Protocol 2.20, 2.23
This version utilizes a bit upgraded header information:
<SOH>BD<STX>SNO<HT>152<LF>DATE<HT>19980715<LF>TIME<HT>114500<LF>
SID<HT>2<LF>PID<HT>26<LF>NAME<HT>JOE
SMITH<LF>MODE<HT>0<LF>WRN<HT>0<LF>
PM1<HT>12<LF>PM2<HT>204<LF>RM1<HT>51<LF>WM1<HT>23<LF>WM2<HT>57<LF>
WM3<HT>92<LF>PARN<HT>22<LF>P01<HT> 6.6<HT>0<LF>P02<HT>4.29<HT>0<LF>
P03<HT> 167<HT>0<LF> ... P21<HT> 8.2<HT>0<LF>P22<HT> 5.3<HT>0<LF>
AGE<HT>24<LF><ETX>7C<EOT>
A parameter called AGE is included in the header info. This represents the age of the given
patient (if selected in service menu, and defined at sample info screen).
The interpretation of this value is:
If the age value is greater than 128, then it represents months in the following way:
MONTHS = AGE – 128 (months).
If it is less than 128, then it represents years.
This parameter has been removed in version 2.23
rev 1.09