Installing the Thermometer

If the thermometer is to be used in conjunction with a protection jacket and
air purge (recommended), cross-refer to the Installation Guides provided with
those accessories.
Fig. 3 R4 Thermometer Installation Dimensions

3.1 Positioning the Thermometer

A position should be chosen for the thermometer that is accessible for
servicing and where it is not exposed to heat, fumes and water spray etc. The
path between the thermometer and hot surface should be as free as possible
from smoke or water spray and from the intrusion of machinery.
The axis of the thermometer should be approximately at right angles to
the hot surface. However, an angle of up to 45° from the normal is usually
acceptable.