b. When "Off" has been selected, the user will have full access to the clinical screens without
having to authenticate with a 4-digit PIN before accessing.
5.5.5 Skin Temperature Configuration
The Skin Temperature Configuration is an available option to turn "On" or "Off" within the
Facility's Care Unit. This option cannot be changed on a per Monitor basis on a netoworked
floor.
If Skin Temperature is turned "On" for the Facility's Care Unit through the network, then all
networked Monitors' configurations will be synched to the .XML file once the Monitor connects to the
network again.
Note: Only while in Bio Med Mode is it possible to turn the Skin Temperature vital sign "On" or
"Off."
Note: Un-Networked/Standalone Monitors are the only Monitors that are able to have Skin
Temperature turned "On" or "Off" while in Bio Med Mode.
Configuration Options
Skin Temperature
a. When "On" is selected for Skin Temperature and the Chest Sensor is connected to the Wrist
Monitor, the Skin Temperature vital sign measurement will be displayed.
b. When "Off" is selected for Skin Temperature and the Chest Sensor is connected to the Wrist
Monitor, the Skin Temperature vital sign will not be displayed.
Option Values
On (default)
Off
Level Updateable
Facility: yes
Care Unit: yes
Monitor: yes
Clinical Configurations
Skin TEMP OFF