Exchanging a Monitor With Low Battery
3. Double bump the new Monitor with the Monitor on the patient's wrist.
- or -
Touch
.
Cancel
If cancelled, the Device Swap will automatically be cancelled.
If the monitors are successfully double bumped, the Exchange process will be initiated. The new
Monitor will display the Waiting Patient Transfer screen and the patient's Monitor will display the
Confirm Device Swap screen.
Note: Once the Exchange process has been initiated, the alarms will be paused on the patient's
current Monitor.
cancelled.
Waiting Patient Transfer
(Instructions on the new Monitor)
4. Touch
Confirm
- or -
Touch
.
Cancel
If cancelled, the Exchange process will automatically be cancelled.
If confirmed, the patient's monitoring session will be transferred from the patient's current
Monitor to the new Monitor.
The new Monitor will display the Instructions screen and the patient's old Monitor will display
the Clean Monitor screen.
138
Alarms will be paused until the process is successfully completed or
to replace the existing patient's Monitor with the new Monitor.
Confirm Device Swap
(Instructions on the patient's Monitor)