Equipment Alerts; Visi Mobile Monitor Alerts - Sotera VISI User Manual

Mobile monitoring system
Table of Contents

Advertisement

Equipment Alerts

ViSi Mobile Monitor Alerts

Display Message
Symbol(s)
AUDIO FAILURE
WIRELESS
RADIO
MONITOR
BATTERY
CRITICAL
MONITOR
BATTERY LOW
UNABLE TO
MONITOR
CONNECT TO
PATIENT
ACCEL
MONITOR
Severity
Cause
Low
Either the microphone or
the speaker on the Monitor
has failed.
Low
The wireless radio in the
Monitor is not transmitting.
Low
Monitor battery charge is
critically low.
Low
Monitor battery charge is
low.
Low
Battery in the Monitor is too
low to continue monitoring.
N/A
Low
A sensor has been
connected to the Monitor
but not yet applied to the
patient. No vital sign
measurement has been
detected.
Low
Unexpected error occurred
with the accelerometer in
the Monitor.
Equipment Alerts
Solution
Stop monitoring and replace
the Monitor.
Return the Monitor to Sotera
Wireless, Inc.
Stop monitoring and replace
the Monitor.
Return the Monitor to Sotera
Wireless, Inc.
Replace the Monitor.
Prepare to replace the Monitor.
Replace the Monitor.
Apply the sensor to the patient.
Replace the Monitor and return
it to Sotera Wireless, Inc.
201

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VISI and is the answer not in the manual?

Table of Contents

Save PDF