Instrument Use
Program
165 °C Program
2.5 Setting User Programs
Select program
PRG1
PRG2
PRG1: 100°C
OK
Prog
Program name?
( RG1)
P
Program name?
(11
5
)
OK
Temperature?
100°C
OK
Temperature?
115°C
OK
Time?
60'
115 :
115°C
60'
OK
Prog
2.6 Error Messages
Error Message
"Block is too hot, Please
Wait"
"Init Error"
Table 3 Stored Programs (Continued)
Description
Heats vials for 30, 60, or 120 minutes at 165 °C.
The DRB 200 can accommodate three user programs in storage
locations PRG1, PRG2, and PRG3.
1.
Select PRG1 from the Select Program menu.
60'
2.
Press Prog to enter the programming mode.
OK
3.
Assign a 4-character program name using the appropriate
keys. Press OK to confirm the entry.
4.
Use the up and down arrow keys to set the temperature value
between 37–165 °C. Press OK to confirm the entry.
5.
Use the up and down arrow keys to set the time between
0-480 minutes. Press OK to confirm the entry.
6.
Press OK to confirm the program. The program can be
OK
changed by pressing Prog. Press the up arrow key to return
to the program selection screen.
Problem
The temperature of the heating block is higher than the target
temperature. Wait until the heating block cools.
The instrument is defective. Contact Customer Service.
19