Environmental Requirements - ABB Endura AZ40 Manual

Oxygen and carbon monoxide equivalent coe analyzer
Hide thumbs Also See for Endura AZ40:
Table of Contents

Advertisement

2

Environmental requirements

Humidity
0 to 95 %
55 °C
(131 °F)
Max.
– 20 °C
(–4 °F)
Min.
1500 mm
(59 in.)
(approx.)
Fig. 4 Environmental requirements
Siting – sensor orientation
CAUTION – Damage to equipment
Mount probe horizontally or vertically – perpendicular to
flow direction.
Avoid locations where:
— obstructions or bends create turbulence in the gas flow and /
or hinder probe insertion and removal
— vibration induced by other plant or vortex shedding is present
— the probe may be subject to shock loading, for example,
close to ash hammers, within 3 m (9 ft.) of steam or liquid
process cleaning apparatus
6
IP66 / NEMA 4X
– 20 °C
(–4 °) Min.
70 °C
(158 °F) Max.
–20°C
(–4 °F) Min.
IP55 / NEMA 4
70 °C
(158 °F) Max.
See page 35
for maximum
probe temperatures
and process pressure
– 20 °C
range
(–4 °F) Min.
CI/AZ40-EN Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents