Cardiovascular Systems, Inc. DIAMONDBACK 360 Peripheral Orbital Atherectomy System Exchangeable Series Instructions For Use Manual page 39

Table of Contents

Advertisement

OAS Pump Specifications
Parameter
Depth
Height
Width
Weight
Electrical cable length:
OAS pump to electrical
outlet
Master Fuse
External housing
Electrical connector type
(Main Power)
Audible information
signals
Visual alerts
Storage conditions
Operating conditions
Operating life
Water Ingress Protection
* Timer resets when crown spinning stops.
Value
<30.6 cm (12.0 in)
20.3 cm (8.0 in)
25.4 cm (10.0 in)
<5.0 kg (11 lbs)
6.1 m (20 ft.)
250 V 4A SLOW BLOW (Type T)
ABS Plastic
Mains Power Plug (100–240 V AC @
50–60 Hz)
Audible information signal for
approximately every 25 sec of OAD
spin time.* Audible information signal
every 5 sec for a total of 30 sec when
the saline level falls below 200 mL
during a treatment period.
Start button
Low Saline Information Signal when
≤200 mL (± 100 mL) remaining out of
a 1000-mL bag of saline
Room temperature in a clean
environment.
Typical operating
room/catheterization laboratory
environment (10-30°C)
875 hours minimum, with 350 hours
of therapy minimum or 5 years
IPX1: Protection against water
ingress
34