Caution: Do not use glass bottles for the saline solution with VIPERSLIDE
Lubricant or hang multiple saline bags from the low saline level sensor as this will
disable the Low Saline Level Sensor.
5. Remove the bag spike from the empty bag of saline and Lubricant and spike the
new bag of saline and lubricant.
6. Power on the OAS pump by pressing the green Start button on the OAS pump
control panel.
7. Ensure that no air is present in the saline tubing.
12.1.4.
Replacing the OAD
Caution: Do not reuse cartridges from the existing handle to the new replacement
handle. Replace the entire OAD with both a new handle and a new cartridge.
If the OAD (handle plus cartridge) needs replacing, perform the following:
1. Stop the spinning crown and drive shaft by pressing and releasing the On/Off
button on top of the crown advancer knob.
2. Disconnect the OAD power cord from the OAS pump.
3. Leave the introducer, guide sheath, or guide catheter and the guide wire in place,
release the guide wire brake on the OAD, and retract the OAD sheath and drive
shaft, from the introducer, guide sheath, or guide catheter, while monitoring and
maintaining guide wire position.
4. Power off the OAS pump by pressing the green Start button on the OAS pump
control panel to stop saline from flowing through the saline tubing and verify that
the green LED, on the OAS pump control panel, is not illuminated.
5. Disconnect the saline tubing from the OAD currently in use and set aside for use
with the replacement OAD.
6. Obtain a new replacement OAD and carefully remove the new replacement OAD
from the package.
Note: Removal of the OAD from the packaging by tipping the tray and allowing
OAD to fall out can result in damage to the OAD or ancillary devices underneath
the OAD.
7. First, connect the existing saline tubing to the new replacement OAD, then
connect the new replacement OAD power cord to the OAS pump.
8. Press the green Start button on the OAS pump control panel to start the saline
flowing through the saline tubing and verify that the green LED illuminates.
9. Purge the air from the OAD.