13. Specifications
13.1. OAD Specifications
Parameter
Electrical cable length:
OAD to OAS pump
Electrical connector type
(OAD power)
Fluid connector type
Saline line tubing length
(from pump to OAD port)
Visual alerts
Sterilization
Storage conditions
Operating conditions
Operating life: Cartridge
Operating life: Handle
Water Ingress Protection
Approximate saline flow
rate for 145 cm (4 Fr) OAD
Approximate saline flow
rate for 145 cm (5 and 6
Fr) OAD
Approximate saline flow
rate for 75 cm (4 Fr) OAD
Approximate saline flow
rate for 200 cm (5 Fr) OAD
Approximate saline flow
rate for 180 cm (5 Fr) OAD
Value
3.4 m (11 ft.)
Type CF applied Part –DC barrel
(48 V DC)
Polycarbonate Luer fitting
3.2 m (10.5 ft.) minimum
Speed and brake indicators
Ethylene oxide (EtO) cycle
Room temperature in a clean
environment away from magnets
and sources of electromagnetic
interference (EMI).
Typical operating
room/catheterization laboratory
environment (10-30°C)
8 minutes of total therapy time
24 minutes of total therapy time
IPX1: Protection against water
ingress
7 mL/min to 16 mL/min
10 mL/min to 30 mL/min
10 mL/min to 23 mL/min
10 mL/min to 34 mL/min
16 mL/min to 35 mL/min
33