ENGLISH
CHAPTER 5:
Scanner Handling, Cleaning, and
Disinfection Instructions
HANDLING OF THE SCANNING UNIT (WAND)
The Scanning Unit contains delicate components and should be handled with care.
•
HANDLING OF THE SCANNING UNIT CABLE
The scanner cable should be treated with care to avoid possible damage.
•
Avoid twisting, knotting, pulling or stepping on cable, etc.
•
Between patient sessions, it is recommended to undo any twists and knots in order to
•
relieve all tension from the scanner cable
RECOMENDED BEST PRACTICES FOR CLEANING AND DISINFECTING
the Scanning Unit, Base Unit, Wheel Stand and/or Counter Stand in between patients.
• Do not spray disinfectant directly on scanner system surfaces.
• Spray the disinfectant on a towel, or use disinfectant wipes for the Scanning Unit,
and Base Unit.
• Warning: over saturation of disinfectant product on the scanner system surfaces may
cause damage, including internal components.
• Follow the disinfectant manufacturers' instructions for appropriate contact time.
Remove residual liquid disinfectant with a lint-free, clean cloth.
• Note: follow standard precautions for personal protection, as appropriate.
• Warning: DO NOT touch the optical surface of the Scanning Unit (Wand).