Maintenance And Cleaning; Batteries - Nedap V-Scan User Manual

Table of Contents

Advertisement

4. Maintenance and cleaning

4.1. Batteries

The V-Scan is powered by four rechargeable batteries. The full performance of new batteries is
achieved only after a few charge and discharge cycles. The ambient temperature for charging
the batteries of the V-Scan must be between 5°C and 25°C. Only use the charger ( 230/110V
AC – 12VDC) provided along with the V-Scan.
Before first use of the V-Scan, charge batteries at least 16 hours for obtaining the
maximum battery capacity.
Connect the plug in the connector on the right side to charge the batteries.
The batteries can be charged and discharged hundreds of times, but it will eventually wear out.
When the scanning time is noticeable shorter than normal, replace the batteries. Warranty is not
applicable for the rechargeable batteries.
Steps for replacement of rechargeable batteries:
1. Open the back cover of the V- Scan. Use
PZ2 screw driver.
2. Open the back cover.
3. Replace the batteries. Mount the batteries
in the correct direction. Watch the positive
and negative terminals of the batteries for
the direction.
Dispose of batteries according to local
regulations (e.g. recycling). Do not dispose
as household waste.
4. After replacing batteries, close the back
cover and fix the screws firmly.
5. Connect the charger, now the V-Scan can
be switched on.
Only replace batteries with same type (4x AA-size, rechargeable, Recyko+, 1.2V,
2100mAh, NiMH, GP210AAHCB)
VP5002-300PM-00GB-V-Scan
Manual version 2.7 / Page 9