Megohmmeter
Hide thumbs Also See for 800:

Advertisement

ACL Incorporated
1960 E. Devon Avenue, Elk Grove Village, IL 60007
To increase the accuracy merely adjust the calibration at the use temperatures and humidities 
the meter will be used.  For example, if you were using the meter at 70º F and 70 RH you would 
calibrate the meter exactly at those conditions using a 1% accuracy standard thermometer and 
relative humidity meter. 
 
The meter will be stable and stay accurate at those conditions. 
 
Allow the meter to equilibrate at those environment conditions. It sometimes takes ½ hour for 
the meter to adjust to new environment conditions.   
 
 
 
O
A
:
PTIONAL 
CCESSORIES
M
P
C
INIATURE 
ROBES WITH 
Two miniature conductive rubber external probes
One shielded BNC coaxial molded cable with gold center pin
Two female banana jacks to fit an ACL 800C meter
Made in USA
Will measure resistance of all small surfaces according to ESD S11.11
Replaceable probes
Conforms to all EOS and CE test procedures
M
R
EASURING 
ESISTANCE TO 
Measure resistance of ESD garments
Stainless steel surface areas
Test according to ESD-STM2.1-1997
Made in USA
One year warranty
MKB June 18, 2008
ORDS
G
(RTG)
ROUND 
G
C
 
ARMENT 
LIPS FOR TESTING 
847-981-9212
ESD
S
 
MOCKS
Page 12 of 12