Setting the test gas concentration

Menu path
Test gas concentration
EL3060 CONTINUOUS GAS ANALYZERS
Setup
Calibration Data
The set points of the test gas concentrations for the zero-point and the
end-point calibration must be set separately for the manual and automatic
calibration.
The set points of the test gas concentrations for the manual calibration can
also be set during calibration (see page 68) and in the configurator (see
page 63).
Zero-point set point
Value range: Lower range value of the physical measuring range minus 20 %
of the measurement span up to the upper range value of the physical meas-
uring range
End-point set point
Value range: Lower range value of the physical measuring range up to the
upper range value of the physical measuring range plus 20 % of the meas-
urement span
|
OI/EL3060-EN REV. A
Test Gas Set Points
61