Error message
188
_CP_LCM_Manual_EN-UK final.indd 188
_CP_LCM_Manual_EN-UK final.indd 188
Meaning
You drew up blood or control solution
too soon with the test strip, i.e. before
the test strip symbol and drop symbol
were flashing in the display.
Turn the meter off and then on again
by pressing the Test button. Repeat
the test with a new test strip. Only
draw up blood or control solution with
the test strip when the test strip sym-
bol and drop symbol are flashing in
the display.
You bent or moved the test strip during
the test either while you were drawing
up blood or control solution with it or
during the test.
Turn the meter off and then on again
by pressing the Test button. Repeat
the test with a new test strip.
Symbols, error messages and troubleshooting
20.6.2007 11:04:45
20.6.2007 11:04:45