4. Replace the base fi rmly with pressure while tightening the screw. The meter will turn on automatically. 5. Tighten the screw with #4 Phillips head screwdriver. SCREW Be sure not to over tighten. Unscrew to remove base Sensafe.com...
FILL, DIP, READ TOTAL BROMINE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Page 5
Place the Cell Cover onto the CELL and result displayed as Total Bromine. This result is automatically stored in the bR menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ CHLORINE DIOXIDE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Place the Cell Cover onto the CELL and result displayed as Chlorine Dioxide. This result is automatically stored in the Cd menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ FREE CHLORINE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
5-7 using same sample and a fresh strip. This result is automatically stored in the CL menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ TOTAL CHLORINE HIGH POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Page 11
Place the Cell Cover onto the CELL and result displayed as Total Chlorine. This result is automatically stored in the CLH menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ HYDROGEN PEROXIDE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Page 13
Place the Cell Cover onto the CELL and result displayed as Hydrogen Peroxide. This result is automatically stored in the HP menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ OZONE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Page 15
Glycine strip into the reacted sample from step 7. Take this second result and subtract it from the fi rst result (obtained in step 7 above). This new value is your Ozone result. For a more detailed procedure, visit sensafe.com/exact-ez-chlorine-plus. Sensafe.com...
FILL, DIP, READ PERACETIC ACID POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
Page 17
Place the Cell Cover onto the CELL and result displayed as Peracetic Acid. This result is automatically stored in the PA menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
Page 18
FILL, DIP, READ Requires water sample with a Total Alkalinity minimum of 20ppm. POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to...
Page 19
Place the Cell Cover onto the CELL and result displayed as pH. This result is automatically stored in the PH menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
FILL, DIP, READ TOTAL CHLORINE POWER ON PHOTOMETER Press the button to power on the Chlorine eXact ® FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with FILL the water sample to be tested, then cell to capacity to begin test.
5-7 using same sample and a fresh strip. This result is automatically stored in the TC menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall. Sensafe.com...
ACRYLIC CALIBRATION KEY The Acrylic Calibration Key is a tool for verifying the calibration of your Chlorine eXact EZ Photometer ® compared to its original factory settings. Run this test immediately upon receipt and record value for future reference. Follow steps 1-4 on page 8 (do not use cap during ZERO).
If water temperature is above 95°F/35°C, press READ when the timer displays “10” for Total Bromine, Free Chlorine, Total Chlorine, and pH tests. Cleaning the cell with water and brush after each test is recommended for best accuracy and prior to storage of unit. Sensafe.com...
TROUBLESHOOTING/INTERFERENCES Listed below are possible situations that may arise while testing. Please contact one of our knowledgeable customer service representatives if you require further assistance. Subject Cause Solution Dim screen or no Low battery Replace batteries response from meter “LO” on LCD while Low battery Replace batteries zeroing...
If the meter is faulty or otherwise defective contact ITS by phone (+1-803-329- 9712 Ext. 0) or email (its@sensafe.com) to describe the problem and obtain a return authorization form before returning the photometer to ITS. Damage caused by improper packing of the photometer for return shipment to ITS will not be covered by the warranty.
Each box contains 10 ampoules of 10mL solution with no dilution necessary. For Ready Snap 3 values, visit us online at ® sensafe.com/exact-ez-chlorine-plus/ COMPLIANCE TESTING This DPD test system for Chlorine and Chlorine Dioxide are accepted for reporting by most health departments because the...
For additional technical support, call (803) 329-9712. Industrial Test Systems, Inc. 1875 Langston Street, Rock Hill, SC 29730 800-861-9712 or 803-329-9712 its@sensafe.com | sensafe.com ITS Europe, Ltd. The UK Centre for Homeland Security Building 7, Chilmark, Salisbury, Wiltshire SP3 5DU UK +44 1722 717911 Manufacturing quality products since 1989 itseurope@sensafe.com | itseurope.co.uk...
Value provided represents best possible accuracy under laboratory conditions, but may vary throughout the detection † range. For a complete list of accuracies throughout all ranges, please visit sensafe.com/exact-ez-chlorine-plus/. Requires the use of 2 strips if reading is above 6 ppm.