Shock and Vibration

To meet these specifications, you must panel mount the system and
either affix ferrules to the ends of the terminal wires or use the
NI 9971 backshell kit to protect the connections.
Operating vibration
Random (IEC 60068-2-64)......... 5 g
Sinusoidal (IEC 60068-2-6) ....... 5 g, 10 to 500 Hz
Operating shock
(IEC 60068-2-27).............................. 30 g, 11 ms half sine,

Electromagnetic Compatibility

This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement,
control, and laboratory use:
EN 61326 EMC requirements; Industrial Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
NI 9225 Operating Instructions and Specifications | © National Instruments | 27
, 10 to 500 Hz
rms
50 g, 3 ms half sine,
18 shocks at 6 orientations