NeurOptics NPi-200 Instructions For Use Manual page 8

Pupillometer
Hide thumbs Also See for NPi-200:
Table of Contents

Advertisement

bottom of the NPi-200 Pupillometer handle, and the gold connector pins located in the base of the NPi-200
Charging Station.
DO NOT
use an oversaturated cloth. Be sure to squeeze out excess liquid prior to wiping the NPi-200
Pupillometer or the NPi-200 Charging Station.
DO NOT
allow the cleaner to collect on the instrument.
DO NOT
use any hard, abrasive or pointed objects to clean any part of the NPi-200 Pupillometer or NPi-200
Charging Station.
DO NOT
immerse the NPi-200 Pupillometer or the NPi-200 Charging Station in liquid, or attempt to sterilize the
product, as damage to the electronic and optical componentry could occur.
Drying and Inspection Following Cleaning
Confirm the NPi-200 Pupillometer is thoroughly dry before placing in the NPi-200 Charging Station to charge. Once
thoroughly dry, place the NPi-200 Pupillometer into the NPi-200 Charging Station and plug in the NPi-200 Power
Supply to the back of the charging station to power ON.
DO NOT
place the NPi-200 Pupillometer into the NPi-200 Charging Station until all components are
completely dry.
DO NOT
reconnect the NPi-200 Power Supply to the NPi-200 Charging Station until all components are
completely dry.
Cleaning Considerations: Gold Connector Pins and Blades
In instances where there is concern of exposure to highly resistant bacteria or viruses (ie: Clostridium difficile, or "C.
diff"), we understand that hospital protocols may require use of cleaning solutions containing sodium hypochlorite
(bleach) when cleaning equipment. Please be aware solutions containing sodium hypochlorite (bleach) will corrode
the gold connector blades located on the bottom of the NPi-200 Pupillometer handle (Figure 1), and the gold
connector pins located in the base of the NPi-200 Charging Station (Figure 2.)
NPi®-200 Pupillometer
Figure 1- Gold Connector Blades
DO NOT
use products containing sodium hypochlorite (bleach) to clean the gold connector blades located
on the bottom of the NPi-200 Pupillometer handle, and the gold connector pins located in the base of the
NPi-200 Charging Station.
NeurOptics® NPi®-200 Pupillometer System—Instructions for Use © 2018 NeurOptics, Inc.
NPi®-200 Charging Station
Gold
Connector
Blades
Figure 2- Gold Connector Pins
Gold
Connector
Pins
8