Quick Links

BabyWeigh
II Scale
Top
Tray Locking
Lever
Display Panel
Operation Keys
stabilized
net
lock
before
after
amount
feed
feed
fed
Bottom
Battery Cover
AC Adapter
Inlet
RS-232C
Connector
Medela, Inc., 1101 Corporate Dr., McHenry, IL 60050
Phone: 800-435-8316 Fax: 815-363-1246 Email: customer.service@medela.com www.medela.com
BabyWeigh and #1 Choice of Hospitals and Mothers are trademarks and Medela is a registered trademark of Medela Holding AG.
© 2013 Medela, Inc. 1908171 B 0313
Baby Tray
(for infant)
Level Gauge
weight
reweigh
units
tare/zero
off
lock off
fed baby
g / lb. oz
3
2
1
ON
amount
after
fed
before
feed
feed
reset
Adjusters (4)
BabyWeigh
II Scale

Operating Instructions

Considerations to follow prior to test-weighing:
Diaper should not be changed between the "before" and "after" weights.
Baby should wear the same clothing for "before" and "after" weights.
Use a bib, but do NOT include in "before" and "after" weights.
To improve accuracy, bundle active infants and verify that all blankets, clothing, and extremities are
inside the basket and not touching the side of the
Record weights in grams only.*
*Note – Grams is closely equivalent to milliliters (mL) or cubic centimeters (cc). Meier PP, et al. A New Scale for in-Home Test-Weighing for Mothers
of Preterm and High Risk Infants. Journal of Human Lactation, 1994;10(3); 163-168.
Instructions for Weighing
Setup:
1. Attach the tray and level the scale prior to weighing.
2. Press the
button and wait until a '0' appears in the display. If using a blanket or pad on
the scale, press
tare/zero
button to zero out the excess weight. Net
3. Level scale using bubble gauge and adjusters on the bottom of the scale.
Before-feed Weight (pre-weight):
1. Place baby on the scale, holding one hand just above baby. Wait until
the display indicates lock
(Beep sound). The value that appears is the
pre-weight.
1
2. Press the
button to store the pre-weight.
3. Remove baby from the scale, place baby in a safe location and then
record pre-weight. The display will reset to '0' within a few seconds.
The pre-weight will be stored in memory.
Feed baby. Leave scale on during feeding.
After-feed Weight (post-weight):
1. Ensure the display has reset to '0'. If it has not, press the
to zero out the display.
scale.
will be indicated.
button

Advertisement

loading
Need help?

Need help?

Do you have a question about the BabyWeigh II and is the answer not in the manual?