7 TROUBLESHOOTING/MAINTENANCE - USER/DEALER
SYMPTOM
• Bed section does not move.
• Bed section does not stop
moving.
• Bed Hi/Lo movement does not
stop.
®
Invacare
G5510 Bed
PROBABLE CAUSE
• Broken connector cap.
• Motors are not functioning.
• Hi/Lo Motor not
functioning.
• Thermal Protection
Activated.
• Pendant.
• Defective Pendant.
• Damaged Cables.
USER SOLUTIONS
• Call dealer.
• Call dealer.
• Ensure cables are not entangled or
pinched.
• Call dealer.
• Allow bed unit to sit for up to 30
minutes and retry.
• Check for depressed/stuck
buttons. If button is stuck, call
dealer for repair. Unplug power
cord from the wall.
• Check cables, if damaged call
dealer.
60
DEALER SOLUTIONS
• Call Invacare for instructions to
replace connector cap.
• Check to make sure nothing is
blocking actuator.
• Ensure cables are not entangled or
pinched.
• Ensure Drive shaft is correctly
installed.
• Ensure Hi/Lo motor cable is
correctly connected. Refer to
Assembling the Drive Shaft on
page 51.
• Allow bed unit to sit for 30
minutes and retry.
• Replace pendant. Refer to
Electrical Connections on page 55.
• Check cables, if damaged, replace.
Refer to Assembling the Drive
Shaft on page 51.
Part No 1180159