Sensor length - Compact version
Ʋ Alloy C22 (2.4602)
Ʋ Inconel 718
Sensor length (L) - Tube version
Ʋ 316L, Alloy C22 (2.4602)
Ʋ Inconel 718
Weight
Ʋ Instrument weight (depending on
process fitting)
Ʋ Tube extension
Surface quality
Process fittings
Ʋ Pipe thread, cylindrical (DIN 3852-A)
Ʋ Pipe thread, conical (ASME B1.20.1)
Ʋ Flanges
Max. torque - process fitting
Ʋ Thread G1, 1 NPT
Output variable
Output
Switching voltage
Ʋ Min.
Ʋ Max.
Switching current
Ʋ Min.
Ʋ Max.
Breaking capacity
Ʋ Min.
Ʋ Max.
Contact material (relay contacts)
Modes (switchable)
Ʋ Max.
Ʋ Min.
Accuracy (according to DIN EN 60770-1)
Reference conditions and actuating variables according to DIN EN 61298-1
Ʋ Ambient temperature
Depending on the mounting boss of the vessel.
2)
30
74 mm (2.91 in)
74 mm (2.91 in)
260 ... 3000 mm (10.24 ... 118.1 in)
260 ... 3000 mm (10.24 ... 118.1 in)
approx. 0.8 ... 4 kg (0.18 ... 8.82 lbs)
approx. 1100 g/m (11.8 oz/ft)
R
approx. 3 µm (1.18
a
G1
1 NPT
DIN EN from DN 50, ASME from 1½"
max. 285 Nm (210 lbf ft)
Relay output (2 x SPDT), 2 floating spdts
10 mV
253 V AC, 60 V DC
10 µA
5 A AC, 1 A DC
50 mW
1250 VA AC, 40 W DC
If inductive loads or stronger currents are switched
through, the gold plating on the relay contact surface will
be permanently damaged. The contact is then no longer
suitable for switching low-level signal circuits.
AgNi (Au plated) or AgSnO (Au plated)
Max. detection or overflow/overfill protection
Min. detection or dry run protection
+18 ... +30 °C (+64 ... +86 °F)