Gossen MetraWatt AT3-III E Operating Instructions Manual
Test adapter, for active and passive testing single and 3-phase
electric devices and extension cables in combination
with secutest.../seculife st test instruments
as well as secutest base(10) and pro (feature i01)
Test adapter for the active and passive testing of single and 3-phase electric devices and extension cables in combination with secutest.../seculife st test instruments (m7050…, m7010... and m6930...) (14 pages)
Adapter for flexible ac current sensors geohm flex 1-3 (16 pages)
Summary of Contents for Gossen MetraWatt AT3-III E
Page 1
Operating Instructions AT3-III E Test Adapter For active and passive Testing Single and 3-Phase Electric Devices and Extension Cables in Combination with SECUTEST.../SECULIFE ST Test Instruments 3-349-155-37 as well as SECUTEST BASE(10) and PRO (Feature I01) 13/6.16...
Page 2
SECUTEST Test Adapter AT3-III E Device Plugs and Outlets Self-Test 1 CEE outlet 11 Test key for triggering I = residual current) Δn Δn 3P+N+PE, 32 A, 400 V Connectors for SECUTEST.../SECULIFE ST (as of version 2 CEE outlet 3.1) Test Instrument 3P+N+PE, 16 A, 400 V 12 Earthing contact outlet for power supply to SECUTEST.../...
3-phase electrical devices and extension cables in combination with SECUTEST.../SECULIFE ST (as Note! of version 3.1) test instrument. The AT3-III E test adapter is equipped with Suitable instruments are identified in the following electronic error monitoring which pages with the following abbreviated designation: disconnects the device under test from the SECUTEST.../SECULIFE ST.
(1 – 4) are only connected to the mains test adapter is defective. If this is the case, protective conductor when the SECUTEST.../ the AT3-III E must be removed from service SECULIFE ST test instrument has been set for and repaired before it is used again.
SECULIFE ST into the earthing contact outlet (12) Conductor Resistance → Zero Balancing. at the AT3-III E. If the AT3-III E is out of use for a prolonged period b) Insert the earthing contact plug (13) from the of time, the earthing contact faces of the plug-...
Page 6
SECUTEST test adapter AT3-III E 230/400 V 50 Hz SECUTEST COM V test adapter AT3-III E SECUTEST PRO 230/400 V 50 Hz Figure 1 Connecting Single and 3-Phase Devices to the AT3-III E and the SECUTEST... GMC-I Messtechnik GmbH...
Testing of Differential Current to 4) of the associated plug at the AT3-III E. In during Automatic Testing per Test Standard order to test, only connect the plug and socket at the ends of the respective extension cable to be...
The actual defect must be sion cables, AT3-III E must be connected determined. with line voltage. 230 V 50 Hz SECUTEST SECUTEST 230/400 V 50 Hz Figure 2 Connecting Single or 3-Phase Extension Cables to the AT3-III E with the SECUTEST... GMC-I Messtechnik GmbH...
Page 9
230 V 50 Hz SECUTEST SECUTEST BASE(10)/PRO SECUTEST SECUTEST BASE(10)/PRO 230/400 V 50 Hz Figure 3 Connecting Single or 3-Phase Extension Cables to the AT3-III E with the SECUTEST BASE(10)/PRO GMC-I Messtechnik GmbH...
Contact protective conductor potential at the 0 °C … +40 °C Temperature mains system which is supplying power to the AT3-III E with the test probe at the end of the Line Voltage probe cable from the SECUTEST.../ Waveshape sinusoidal SECULIFE ST, e.g.
Electromagnetic Compatibility These devices may not be disposed with the trash. Please contact our service department Intrinsic Uncertainty EN 61326-1:2006 class B regarding the return of old devices. Intrinsic Uncertainty EN 61326:2006 Mechanical Design Repair and Replacement Parts Service, Protection case: IP40 terminals: IP20 Calibration Center and Rental Instrument Service...
Page 12
Edited in Germany • Subject to change without notice • A pdf version is available on the internet Phone +49 911 8602-111 GMC-I Messtechnik GmbH +49 911 8602-777 Südwestpark 15 E-Mail info@gossenmetrawatt.com 90449 Nürnberg • Germany www.gossenmetrawatt.com...