Sensitivity Switch Calibration - HP 333A Operating And Service Manual

Distortion analyzer
Hide thumbs Also See for 333A:
Table of Contents

Advertisement

Model 333A/334A
SENSITIVITY
Switch
Position 4
Position 3
Position 2
Position 1
C.
d.
e.
f.
g*
h.
TEST OSCILLATOR
ADJUST
OUTPUT (Increase)
FOR 9 mV
+lo
dB
ac4
+20
dB
s2c3
+30
dB
s2c2
+40 dB
ac1
SENSITIVITYVERNlER..
. . . . .
MIN.
SENSITIVITY.
. .
.1 step counterclochise
METER RANGE
.
,
.
.
,
. .
SET LEVEL
MODE..
. . . . . . . . . . . . .
MANUAL
FREQUENCY RANGE
. . . . . . . .
X1K
Frequency Dial.
. . . . . . . . . . . .
60
Turn slot
on
trimmer capacitor C3 parallel
with side casting of instrument.
Set test oscillator frequency to 60 kHz, and
adjust amplitude for indication
of
+2
dB
on
333A/334A meter.
Set 333A/334A COARSE and
FINE BALANCE
controls
to
center of range.
Set 333A/334A FUNCTION switch to
DIS-
TORTION.
Adjust
test
oscillator frequency vernier con-
trol and C3 for definite null indication on
meter of 333A/334A.
Reduce METER
RANGE setting
as
necessary to maintain
on-scale reading.
Observe electronic counter indication. If
test
oscillator frequency is not between
57
kHz
and 63
kHz,
333A/334A frequency
dial is
out o f
tolerance. Follow steps be-
low to bring frequency
d i a l
within tolerance:
Set 333A/334A controls
as in
step b.
Set test oscillator frequency for elec-
tronic counter indication of 60.00
H I ,
and adjust amplitude for indication of
+2
dB
on meter
of
333A/334A.
Set 333A/334A COARSE and FINE
BAL-
ANCE
controls to center of range.
Adjust 333A/334A frequency dial vernier
control for definite null.
Remove knob from frequency dial. Do
not change tuning capacitor, C4, setting
Loosen retaining screws in frequency
dial plate, and slip dial until 60
i s dir-
ectly under mark on indicator. Be
careful not to change C4 setting.
Tighten retaining screws and replace
knob.
5-30.
SENSITIVITY SWITCH CALIBRATION.
a .
Connect test oscillator to 333A/334A
ae
shown
in
Figure 5-1.
NOTE
Selector positions for the SENSI-
T M T Y selector will
be
referred to
as
follows: MIN = Position 1, next
step
=
Position 2, etc., to MAX
=
Position
6.
b.
C.
d.
e.
f.
8.
Section V
Set Distortion Analyzer controls as follows:
FUNCTION..
. . . . . . . .
SETLEVEL
METER RANGE
. . . . . . .
0.01 VOLTS
SENSITIVITY
. . . . . . . . . .
Position 5
SENSITIVITY VERNIER
. . . . . . .
MAX.
Set test oscillator frequency
to
400 Hz,
ad-
just amplitude for indication of 9 mV on
333A/334A meter.
Set
a
reference on meter of test oscillator
and use amplitude control to maintain set
reference whenever frequency of oscillator
is varied.
Set test oscillator frequency to 100 kHz.
Adjust S2C5 for 333A/334A meter indication
of 9 mV.
Set 333A/334A SENSITIVITY switch to each
position listed in Table 5-6, and increase
test oscillator output by 10
dB
f o r each re-
spective decrease
in
SENSITIVlTY switch
position.
Adjust appropriate capacitor for
333A/334A meter indication of 9 mV
a t
each
setting.
NOTE
When switching the SENSITMTY
positions,
be
careful not to move
SENSITIVITY VERNIER control
from
MAX.
position.
5-31. ADJUSTMENT OF FACTORY SELECTED
COMPONENTS.
5-32. Certain components in the Model 333A/334A
are
individually selected to compensate for
va.rying
circuit parameters. These components are designated
l q r an asterisk
(*) on the schematic and in the
parts
list, and the typical
value i s
shown
describes the function of the factory selected compo-
nents
and lists their value
ranges.
Table
5-7
5-33, T R O U BLES
H
0 OTI N
G
PR 0 CEDU R E t .
5-34. This section contains procedures designed to
assist
in the isolation of malfunctions,
These pro-
cedures are based on
a
systematic analysis
of
the
instrument circuits in an effort to localize the prob-
Table
5-6.
Sensitivity Switch Calibration
5-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

334a

Table of Contents