Surface Sampling Device
 
 
GENERAL DESCRIPTION 
The LIGHTNING MVP ICON Surface Sampling Device is a self‐
contained device for use with the LIGHTNING MVP ICON™ or 
® 
LIGHTNING MVP
systems for quality and HACCP‐related 
monitoring of the environment. The MVP ICON and 
LIGHTNING MVP systems detect food and microbiological 
residues by measuring Adenosine triphosphate (ATP) using 
luciferase bioluminescence. ATP is found in most food 
residues and all bacteria, yeast and mold cells.   
Part No. 64003‐100 (100 tests) 
KIT COMPONENTS 
One box of 100 surface sampling devices 
IMPORTANT USAGE GUIDELINES 
It is critical to insert the sampling device 
into the MVP ICON with the two flat 
sides of the sampling device's base 
chamber facing left and right. Improper 
sampling device orientation may result in 
damage to the instrument. 
Do not insert wet or contaminated 
sampling devices into the instrument. Wipe condensate from 
exterior of device before use with a clean, lint‐free cloth. 
Avoid handling the device with wet or contaminated hands. 
Always remove sampling devices from instrument after 
reading. 
Do not store the MVP ICON or MVP instrument on its side or 
back. Avoid storing in humid areas where water leaks are 
possible. 
TEST PROCEDURE 
A. Collecting Sample 
(a) Twist to remove sampling device from body tube. The 
swab tip must be moist. If fully saturated, any excess 
wetting agent should be blotted off by gently pressing 
the swab tip against the interior of the body tube. Be 
careful not to touch the swab or inside the sampling 
device with fingers. 
(b) Thoroughly swab a standard 10 x 10 cm (4 x 4 in) area of 
interest for a typical flat surface. For irregularly shaped 
surfaces, ensure swabbing technique remains consistent 
each time you swab. Reinsert the swab into the body 
tube using a twisting motion to ensure it is sealed.  
Note:  After swabbing, the sampling device is stable for one 
hour prior to activation. 
 
 
 
Flat Side
 
B. Activating Device and Reading Results
(a) To activate the sampling device, hold upright and push 
the plunger down until it is fully depressed. The reagents 
will be dispensed through the sampling device washing 
the sample off the swab tip and into the collection tube. 
(b)  Insert the sampling device into the MVP ICON or MVP 
and read the results. When inserting into the MVP ICON, 
ensure the flat sides of the sampling device base 
chamber face left and right in the instrument. 
Note: Results should be read within two minutes of 
activation. Please refer to the MVP ICON User Guide or MVP 
System Guide for operating instructions. 
INTERPRETING RESULTS 
(a) When using default thresholds in the MVP ICON or MVP 
systems, readings of 2.5 Zones or less indicate the 
surface is considered clean.  
(b) Readings between 2.6 ‐ 3.0 Zones indicate a warning the 
surface is not adequately clean.  
(c) If the reading is greater than 3.0 Zones, the surface is 
considered dirty and should be re‐cleaned and re‐tested. 
CONTROLS 
A. Positive Control 
It is recommended LIGHTNING MVP ICON ATP Positive 
Controls (Part No. 64001‐25) be run every other day to verify 
integrity of the reagent system.  
(a) A positive control should yield a result from 3.1 to 4.3 
Zones. See ATP Positive Control Directions for Use for 
additional information. 
B. Negative Control 
A negative control is an unused yet activated LIGHTNING MVP 
ICON Surface Sampling Device. 
(a) Negative control results should read between 1.0 to 2.5 
Zones.  
(b) Readings above 2.5 may indicate contamination of the 
MVP ICON or MVP sampling chamber or sampling device.  
See ATP Positive Control Directions for Use for additional 
information. Contact BioControl Technical Support at 
800.245.0113 for more information. 
PRECAUTIONS 
The appearance of a yellow band on the swab tip is normal 
and will not affect performance. 
If accidental activation prior to sampling occurs, do not use 
device. 
 

Advertisement

loading
Need help?

Need help?

Do you have a question about the LIGHTNING MVP ICON and is the answer not in the manual?

Questions and answers

Summary of Contents for Biocontrol LIGHTNING MVP ICON

  • Page 1 Zones. See ATP Positive Control Directions for Use for  possible.  additional information.  TEST PROCEDURE  B. Negative Control  A. Collecting Sample  A negative control is an unused yet activated LIGHTNING MVP  (a) Twist to remove sampling device from body tube. The  ICON Surface Sampling Device.  swab tip must be moist. If fully saturated, any excess  (a) Negative control results should read between 1.0 to 2.5  wetting agent should be blotted off by gently pressing  Zones.   the swab tip against the interior of the body tube. Be  careful not to touch the swab or inside the sampling  (b) Readings above 2.5 may indicate contamination of the  device with fingers.  MVP ICON or MVP sampling chamber or sampling device.   (b) Thoroughly swab a standard 10 x 10 cm (4 x 4 in) area of  See ATP Positive Control Directions for Use for additional  interest for a typical flat surface. For irregularly shaped  information. Contact BioControl Technical Support at  surfaces, ensure swabbing technique remains consistent  800.245.0113 for more information.  each time you swab. Reinsert the swab into the body  PRECAUTIONS  tube using a twisting motion to ensure it is sealed.   The appearance of a yellow band on the swab tip is normal  Note:  After swabbing, the sampling device is stable for one  and will not affect performance.  hour prior to activation.  If accidental activation prior to sampling occurs, do not use    device.   ...
  • Page 2   Always insert the sampling device into the MVP ICON with    the base chamber’s two flat sides facing left‐to‐right.    Read sampling device within two minutes of activation.    When reading the sampling device, the MVP ICON or MVP    must be kept upright (no greater than 45° angle). See MVP  ICON User Guide or MVP System Guide for additional    information.    STORAGE  Sampling devices should be stored at a temperature between  2 ºC – 30 ºC (36 ˚F – 86 ºF). Protect from direct sunlight.  Unused devices must be resealed in the foil bags. WARRANTY    BioControl Systems, Inc. (BCS) warrants this product to be free from defects  in materials and workmanship, when stored under labeled conditions and  used as intended until the expiration date stated on the package. BCS  agrees during the applicable warranty period to replace all defective  products after return to BCS. BCS shall not have obligation under this  Limited Warranty to make replacements which result, in whole or in part,  from negligence of the Buyer, or from improper use of the products, or use  of the product in a manner for which it was not indicated. Buyer shall notify  BCS of any products which it believes to be defective during the warranty  period. At BCS option, such products shall be returned to BCS,  transportation and insurance prepaid. BCS shall replace any such product  found to be defective, at no charge. Should BCS examination not disclose  any defect covered by the foregoing warranty, BCS shall so advise Buyers  and dispose of the product in accordance with Buyer’s instructions.     ©BioControl Systems, Inc. 2013. Printed in USA. LIGHTNING MVP ICON is a  trademark of BioControl, pending registration. LIGHTNING MVP is a  registered trademark of BioControl. US patents No. 7,399,984; 7,544,961;  7,030,403; 6924498. Australian patents 2002357107; 2003231770. New  Zealand patents 533051, 535819. Japanese patents 4,467,304; 4,431,034. ...