ICA4H Connections
The power supply for the ICA4H is 24 V dc. The ICA4H can operate over the power supply range 13 V to 28 V.
However, the minimum supply voltage is determined by the input impedance of the receiver or monitor connected to
the loop – see the 'Output Shunt Resistance Formula' at the end of this chapter.
Figure 2.7 Connection Details for the ICA4H – Source mode
Figure 2.8 Connection Details for the ICA4H – Sink mode
Figure 2.9 Connection Details for Best EMC Immunity for ICA4H
Take note of the grounding arrangement particularly the bolt hole which capacitively couples the common of the ICA
electronics to the strain elements in the load cell to improve the EMC performance.
6
Mantracourt Electronics Limited – ICA User Manual