Activate the AED Sentinel Monitoring Unit
4.
Check the image and AED status. Use your mobile device or computer to log into
your AED Sentinel Manager account and navigate to the Monitored AED associated
with the AED Sentinel monitoring unit you are installing (based on the monitoring
unit's Sentinel ID). Confirm the image the AED Sentinel monitoring unit just captured
displays the AED's status indicator in the center of the picture and is not obstructed
in any way. Also confirm the AED's status shows "Ready" in AED Sentinel Manager. If
needed, readjust the camera, AED or cables, turn the battery pack off and then on,
and press the button on the communications module again. Repeat the activation
process until you capture an image with the AED's status indicator in the center and
the AED status shows "Ready."
5.
If needed, affix the AED cabinet door mounting system window plate. Check that
the small magnet on the bottom of the camera's mounting system magnetically
attaches to the inside bottom of the cabinet door. If it touches only the glass, peel the
protective backing off of the included window magnet (4a) and adhesively attach the
plate to the outside of the cabinet window opposite the bottom mounting block (4b).
This will magnetically secure the bottom mounting block to the inside of the cabinet
window.
6.
Affix cabinet sticker. Now that you have successfully installed the AED Sentinel
monitoring unit, affix the cabinet sticker to the front of your AED cabinet letting
everyone know AED Sentinel is on watch.
4a)
4b)
5)
CABINET
DOOR
CABINET
STICKER
14
PEEL BACKING
FROM BACK OF
WINDOW PLATE
BOTTOM
MOUNTING
BLOCK
MAGNET
WINDOW PLATE