Advertisement

Quick Links

Installation Guide
AED Sentinel Monitoring Hardware

Advertisement

Table of Contents
loading

Summary of Contents for Readiness Systems AED Sentinel

  • Page 1 Installation Guide AED Sentinel Monitoring Hardware...
  • Page 2 Warranty Readiness Systems warrants that, during the AED Sentinel Monitoring Services subscription term, the AED Sentinel Remote Monitoring Hardware will be free from defects in materials and workmanship. Compliance Statement This device complies with Part 15 of the FCC Rules. Operation is subject to the following...
  • Page 3: Table Of Contents

    This guide walks you through the easy installation process. A typical installation takes no more than 10 minutes. Once installed, AED Sentinel will help ensure your AEDs are always ready when you need them. Note for Non-Metal and...
  • Page 4: Visual Status Indicators And Camera Aiming

    (LCD) icons. AED Sentinel can monitor both status indicator types. AED Sentinel uses a camera to monitor the status indicator on each of your AEDs. When you install AED Sentinel hardware in your AED cabinets, make sure the AED Sentinel camera is properly aimed at each AED’s status indicator.
  • Page 5: Before You Begin

    Follow these steps to install your AED Sentinel monitoring hardware: Drill the antenna hole To work properly, AED Sentinel needs to use the included external antenna to connect to the cellular network. Drill a 9/32”-inch (7.1mm) hole in the top of your AED cabinet to accommodate the cable that connects the external antenna with the AED Sentinel communications module.
  • Page 6: Install The Battery Pack

    Install the AED Sentinel Monitoring Hardware Install the battery pack Insert batteries. Make sure the battery pack is turned off. Slide the cover to remove it and insert all the included batteries, taking care to position them as shown in the battery compartment.
  • Page 7: Install The Cable Management Clip

    Install the AED Sentinel Monitoring Hardware COMMUNICATIONS MODULE LED LIGHT AND ROUND BUTTON FACING TOWARD CABINET DOOR Install the cable management clip Attach the cable management clip. Peel the protective backing off one of the included cable management clips (we’ve provided one extra). Attach it to the ceiling of the cabinet midway between the hinged side of the cabinet and the antenna hole with the open end facing front, toward the cabinet door.
  • Page 8: Place The Aed

    RAIL 180˚ AED Sentinel works regardless of whether your camera lens is in the up or down position. If the camera for this AED is flipped to the UP position, check the box on the Add or Edit Monitored AED page in AED Sentinel Manager to ensure...
  • Page 9: Aim The Camera

    (Instructions for this step are located on page 14 of this installation guide.) Aim the camera For AED Sentinel to monitor your AED, the camera lens must be properly aimed at the AED’s status indicator. The AED cabinet door mounting system allows the camera to be positioned horizontally along the cabinet door and vertically along the rails.
  • Page 10 Install the AED Sentinel Monitoring Hardware Check the camera lens and AED position. Close the cabinet door and, through the glass, view the three aiming notches on the back of the camera to check the position of the camera lens in relation to the AED’s status indicator.
  • Page 11 Install the AED Sentinel Monitoring Hardware Aim vertically. Viewing the status indicator at eye level, use the aiming notches at the sides of the camera to check whether the camera lens points at the center of the status indicator. If it doesn’t, open the door and slide the camera up or down on the rails to adjust its position.
  • Page 12: Activate The Aed Sentinel Monitoring Unit

    Activate the AED Sentinel Monitoring Unit Follow these steps to activate your AED Sentinel monitoring unit: Add your AED to AED Sentinel Manager. Log in to your AED Sentinel Manager account, go to the Monitored AEDs page and click Add Monitored AED. On the Add Monitored AED page, use the Sentinel ID found on the AED Sentinel communications module you just installed (the number is also on the box).
  • Page 13 See Troubleshooting below. The light will continue to blink for about 1 minute. Then … Solid green: The AED Sentinel monitoring unit is completing its activation sequence. The solid green light will stay on for the remainder of the 3-minute AED Sentinel activation period.
  • Page 14 AED’s status indicator in the center of the picture and is not obstructed in any way. Also confirm the AED’s status shows “Ready” in AED Sentinel Manager. If needed, readjust the camera, AED or cables, turn the battery pack off and then on, and press the button on the communications module again.
  • Page 15: Troubleshooting

    AED and send you alerts and messages if the AED cabinet door is opened, the AED status indicator changes from Ready to Not Ready, or if you need to replace your AED batteries or pads. As always, AED Sentinel is here to keep a watchful eye. If you need help, simply contact us at support@aedsentinel.com.
  • Page 16 We’re here to help. aedsentinel.com | support@aedsentinel.com AED Sentinel® is a registered trademark of Readiness Systems, LLC. Patents pending. © 2018 Readiness Systems, LLC – All rights reserved Version 2018.2...

Table of Contents