wtw TecnoLine LF 171 Operating Manual page 93

Conductivity industrial monitors
Table of Contents

Advertisement

Available languages

Available languages

Parameterization
Notes on setting:
When setting the control values the display shows:
l the selected parameter
l the control value (conductivity, temperature, hysteresis) and
the delay time DL before a relay closes or opens
l the relay number
Setting:
The set hysteresis works symmetrically with positive and negative deviation around the set control
value.
Example: With an upper control value UL of 2.85 mS/cm and a hysteresis of 0.15 mS/cm
the relay switches when the measured value exceeds 3.00 mS/cm and switches back
when the measured value goes below 2.7 mS/cm.
in the message line
in the upper 7 segment display.
in the lower 7 segment display.
Submenu: "PARA UL/LL".
With
enter submenu
"PARA UL/LL".
With
select PARA UL LF
With
enter selected submenu: Setting of
upper limit value UL LF (relay 1).
<- Relay number (e.g. 1) is displayed.
With
set the flashing figure
(setting range 0-9).
With
accept and go on to next figure.
Here the maximum setting range is the
configured recorder range Rec 1.
After the last figure go on to HYST OG LF.
With
set the flashing figure and enter
with
.
Setting range is 0 ... ±10 % of recorder
range.
After the last figure accept with
to "DL UL LF".
and go on
97