Control Panel Functions
Time Adjustment and Timer Operation:
and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run time.
Adjustments may be made prior to a run or after a run begins. A quick tap will adjust the time by 30
seconds. Hold down the button for 1 minute adjustments - continue to hold for 5 minute adjustments.
Press the down button once while 00:30 is displayed to access the "hold" feature. When a "hold" run
is initiated, the timer will count up to keep track of total run time. The unit will run for a maximum of 99
minutes and 30 seconds and will then shut off automatically. The "hold" run may also be terminated
by pressing the STOP/OPEN button. The total run time during a "hold" run is displayed on the time
display until the operator unlocks the lid by pressing the OPEN button or opens the lid (during the
automatic unlock after the run ends).

Speed Adjustment / RCF Display:

display to change to run speed. Adjustments may be made prior to a run or after a run begins. A quick
tap will adjust the speed by 50 RPMs. Hold down the button for 100 RPM adjustments. Continue to
hold for 500 RPM adjustments.
Press the RCF button at any time to convert the current speed setting to the corresponding force
value (xg). The speed may be set by RCF by holding down the RCF button while using the up/down
speed set buttons. The jumps made in the RCF value while holding down the RCF button and pressing
the up and down arrow buttons correspond to 50 RPM adjustments in speed. Since the RCF is a
function of the square of the speed, this jump size will vary.

Starting and Stopping a Run:

a run. Press the STOP/OPEN button at any time during a run to terminate it.

Unlocking the Lid:

The lid unlocks automatically for 60 seconds at the end of a run. The lid can also be unlocked for an
additional 15 seconds by pressing the OPEN/STOP button while the unit is idle. The lid cannot be
unlocked while the rotor is spinning.
Imbalance Detection Reset:
is detected, the centrifuge will terminate the current run and will begin to brake to a stop. The words
"BALANC" and "ERROR" will flash on the speed display. Once the rotor has stopped, open the lid
the cancel the error reporting, balance the load and begin a new run. Alternately, the error reporting
can be canceled by pressing the OPEN/STOP button

Advanced User Settings

This centrifuge allow for total control of the run by the user. You may adjust the acceleration rate,
adjust the braking rate, adjust the imbalance sensitivity level, select from several audible settings,
and use the optional countdown delay.

Acceleration Rate:

using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until "ACCEL" is displayed in the
speed display.
3. The current acceleration rate is displayed in the time display. Use the 'UP/ DOWN arrow buttons
next to the time display to adjust the acceleration rate from 1 (minimum) to 10 (maximum).
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display
to adjust additional settings.
10
The run time may be set from 30 seconds to 99 minutes
Press the up and down arrow buttons next to the speed
With the lid switch closed, press the START button to begin
The lid is unlocked whenever the red UNLOCKED indicator light is illuminated.
This centrifuge is equipped with imbalance detection. If an imbalance
The user may adjust the acceleration rate from minimum (1) to maximum (10)