TESTING AND CARE
Testing
Visual
Inspection
Operation
Quick Touch Pendant User Guide
0510-0516-C
The Quick Touch Pendant should be tested
monthly to ensure operation.
CAUTION: You must test all pendants prior
to use to verify proper operation. This
includes every time that the band is
replaced. Failure to test the pendants before
use can result in system failure and/or an
elopement. In addition, failure to test
transmitters voids the RF Technologies
Product Warranty.
1. Visually inspect the pendant for damage
or loose parts.
1. Press the activation button for at least
one second and ensure that the red
LED flashes, the pendant vibrates and
an Assistance Required alarm is
generated.
2. Clear the alarm using the Alarm
Clearance Card (place the dot on the
alarm clearance card over the activation
button.
3. ensure the red LED stops flashing, and
the blue LED rapidly flashes 6 times to
indicate the alarm has been cleared.
Testing and Care
Page 17 of 22