OPERATION

3.1
Measurement and Confi guration
The Easidew PRO I.S. can be confi gured to provide an output of 4-20 mA (2-wire
connection) for the following:
Dew point
Moisture content in gas
Moisture content in liquids
The Easidew PRO I.S. can be purchased factory confi gured as required. Alternatively, the
Easidew PRO I.S. can be confi gured by the customer, using the Easidew Communications
Kit (EPR-CK) and Easidew Application Software. The Easidew Communications Kit can
be purchased from Michell Instruments or a local representative. For a free copy of the
Application Software contact Michell Instruments' UK offi ce (see www.michell.com for
details of Michell's contact information).
For moisture content in gas, the calculation from the measured dew point is assumed to
be at atmospheric pressure. Alternatively, a fi xed gas pressure needs to be programmed
into the Easidew PRO I.S.
For moisture content measurement in liquid, the Easidew PRO I.S. requires the saturation
constant of the liquid to be programmed into the transmitters, either at the factory or
by the customer using the Application Software.
The transmitter requires a 6-point look-up table for saturation constants up to 3000
ppm
over the temperature range 0 to +50°C (+32 to +122°F). Saturation constants
W
for 8 common liquids can be programmed into the Easidew PRO I.S. via the Application
Software. Alternatively, the user can program saturation constants manually. The
Application Software Help fi le provides detailed instructions on how to perform this task.
3.2

Sampling Hints

Operation is very simple, assuming the following installation techniques are adhered to:
Be Sure the Sample is Representative of the Gas Under Test:
The sample point should be as close to the critical measurement point as possible. Also,
never sample from the bottom of a pipe as entrained liquids may be drawn into the
sensing element.
Michell Instruments
-110 to +20°C (-166 to +68°F)
0 - 3000 ppm
V
0 - 3000 ppm
W
OPERATION
9