2.2 Z23-1 shielding tent
The shielding tent has been designed to shield the measurement set-up against external RF fields.
In case of problems in the VHF range or particularly high demands on the unit under test as in the
automobile industry, for instance, a conductive housing has to be used as a shield against ambient
disturbance fields. Shielding in the immediate vicinity of the unit under test is possible because all
cables to the unit under test are disconnected, coupled to the base plate through capacitive
coupling or filtered in a typical ESA1 measurement set-up.
The tent poles are folded upon delivery for ease of transport. Two rubber plugs snap into place in
the two rear openings of the ground plate to secure the tent against displacement (Figure 4).
Figure 4 ZG 23-1 tent poles folded (on GP 23 ground plate)
To put up the shielding tent, unfold the poles and insert them into the two rear openings of the
ground plate (Figure 5).
Figure 5 ZG 23-1 set up on the GP 23 ground plate