Isotech milliK User Maintenance Manual/Handbook page 7

Hide thumbs Also See for milliK:
Table of Contents

Advertisement

The milliK is likely to be connected to thermometer sensors in use and the user should take
care to ensure that the complete system is safe. For example, metal sheathed thermometers
!
may be connected to the milliK and then placed in a furnace powered from a 230V electrical
supply. Single fault conditions in such a furnace could lead to the thermometer wires and the
front terminals of the milliK, to which they are connected, becoming electrically live and
therefore a hazard to the user. Suitable precautions should be taken, such as using an
isolating transformer in the supply to such a furnace. If you require further advice on safety
issues, please contact Isothermal Technology or one of our appointed distributors - we have
extensive experience of thermometry and can provide advice and equipment to help you.
Retain these instructions. Use only as specified in these operating instructions or the intrinsic
protection may be impaired.
Please observe the following safety precautions:
 Do not use your milliK if it is damaged
 Only connect the power supply to an electrical supply that conforms to the
specification given on its rating plate
 This equipment is for indoor use within an ambient temperature range of 0°C to 40°C
with maximum relative humidity of 95%
 This equipment is for use in moderate climates only. NEVER use the equipment in
damp or wet conditions
 Avoid excessive heat, humidity, dust & vibration
 Do not place liquid filled containers on the equipment
 Do not use where the equipment (or any associated accessories) may be subjected to
dripping or splashing liquids
 Ensure all cables and wires are routed safely to avoid tripping: also, to avoid sharp
bends and pinches
 Clean only with a damp cloth. Do not wet or allow moisture to penetrate the
equipment. Do not use solvents; see section 6 for details of cleaning procedure
 The product should be subjected to regular in-service inspections as required by local
regulations; a yearly interval is suggested
 Do not apply earth test currents to any terminals nor to the shrouds of the USB,
RS232, or ethernet connectors
 The product is designed to comply with EN 61010-1 and can be flash tested. It is
fitted with radio frequency interference suppressors. Therefore, it is recommended
that only a D.C. test be performed. Performing flash tests repeatedly can damage
insulation
 This equipment contains no user-serviceable parts. Refer all repairs to qualified
service personnel. Contact Isothermal Technology or one of our appointed
distributors for details of approved service outlets
 The power supply has been approved to the following safety standards:
©Isothermal Technology
Page 7 of 127
923 milliK manual Iss. 3.02