the '115Ω Calibration Adjustment' window. Use the
keys to select the two required
parameters in turn and press the
key to enter their values:
Press the
OK
(F4) key to start the calibration adjustment. The milliK will then take a number
of readings of the attached resistance standard, calculate and then apply an appropriate
adjustment. A progress bar appears showing the status of the calibration adjustment:
Once started, the calibration adjustment process cannot be aborted.
5.3.3 Adjusting PRT (460Ω) Resistance Range
Connect a calibrated (nominally) 400 resistor to channel 1. In the 'Calibration Menu'
window, use the
keys to select '460Ω Range' and then press the
key to proceed to
the '460Ω Calibration Adjustment' window and use the same procedure described for
calibrating the 115 range (see section 5.3.2) to adjust this range.
©Isothermal Technology
Page 69 of 127
923 milliK manual Iss. 3.02