the '115Ω Calibration Adjustment' window. Use the
keys to select the two required
parameters in turn and press the
key to enter their values:
Press the
OK
(F4) key to start the calibration adjustment. The milliK will then take a number
of readings of the attached resistance standard, calculate and then apply an appropriate
adjustment. A progress bar appears showing the status of the calibration adjustment:
Once started, the calibration adjustment process cannot be aborted.
5.3.3 Adjusting PRT (460Ω) Resistance Range
Connect a calibrated (nominally) 400 resistor to channel 1. In the 'Calibration Menu'
 
window, use the
keys to select '460Ω Range' and then press the
key to proceed to
the '460Ω Calibration Adjustment' window and use the same procedure described for
calibrating the 115 range (see section 5.3.2) to adjust this range.
©Isothermal Technology
Page 69 of 127
923 milliK manual Iss. 3.02