Amprobe 33XR-A User Manual page 15

Professional digital multimeter
Hide thumbs Also See for 33XR-A:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
1.800.561.8187
AC VOLTS (45 Hz – 500 Hz)
RESISTANCE
Ranges: 400 Ω, 40 kΩ, 4 MΩ
Ranges: 400 mV, 4 V, 40 V, 400 V, 750 V
ac
Resolution: 100 mΩ
Resolution: 100 µV
Accuracy: ±(1.0 % of reading + 4 digits) on
400 Ω, 40 kΩ range ,±(1.2 % of reading +
Accuracy:
±(1.5 % of reading + 4 digits)
4 digits) on 4 MΩ range
±(2.0 % of reading + 4 digits) 200 Hz to
Open circuit volts: 0.5 V dc typical,
(3.0 V dc on 400 Ω range)
500 Hz on 4 V range
Peak hold accuracy:
Overload protection:
±(3.0 % + 60 digits) on 40 V to 750 V
1000 V dc or 750 V ac rms
ranges, 400 mV, 4 V ranges unspecified
CAPACITANCE
Input impedance: 10 MΩ
Ranges: 4 µF, 40 µF, 400 µF, 4000 µF
Overload protection: 400 mV Range:1000 V
Resolution: 0.1 uF
dc / 750 V ac rms (15 seconds)
Accuracy:
Other Ranges: 1000 V dc / 750 V ac rms
±(5.0 % of rdg +10 digits) on 4 uF range
DC CURRENT
±(5.0 % of rdg +5 digits) on 40 uF to
Ranges: 400 µA, 4 mA, 40 mA, 300 mA,
400 uF ranges
10 A
±(5.0 % of rdg +15 digits) on 4000 uF
Resolution: 0.1 µA
range
Accuracy: ±(1.0 % of reading + 1 digit) on
Test voltage: < 3.0 V
400 µA to 300 mA ranges
Test Frequency: 10 Hz
±(2.0 % of reading + 3 digits) on 10 A
Input protection: 0.315 A/1000 V fast blow
range
ceramic fuse 6.3×32 mm on µA/mA input
Burden voltage:
TEMPERATURE
400 μA Range:
1 mV/ 1 μA
Ranges:-20 °C to1000 °C, -4 °F to 1832 °F
4 mA Range:
100 mV/ 1 mA
Resolution: 1 °C, 1 °F
40 mA Range:
12 mV/ 1 mA
Accuracy:
300 mA:
4 mV/ 1 mA
±(2.0 % of rdg +4 °C) -20 °C to 10 °C
10 A:
100 mV/ 1 A
±(1.0 % of rdg +3 °C)10 °C to 200 °C
Input protection: 0.315 A/1000 V fast blow
±(3.0 % of rdg + 2 °C) 200 °C to 1000 °C
ceramic fuse 6.3×32 mm on µA/mA input
±(2.0 % of rdg + 8 °F) -4 °F to 50 °F
10 A/1000 V fast blow ceramic fuse
±(1.0 % of rdg + 6 °F) 50 °F to 400 °F
10×38 mm on 10 A input
±(3.0 % of rdg +4 °F) 400 °F to 1832 °F
10 A Input: 10 A for 4 minutes maximum
Overload protection:
followed by a 12 minute cooling period
1000 V dc or 750 V ac rms
AC CURRENT (45 Hz – 500 Hz)
FREQUENCY (autoranging)
Ranges: 400 µA, 4 mA, 40 mA, 300mA, 10A
Range: 4 k, 40 k, 400 k, 4 M, 1 MHz
Resolution: 0.1 µA
Resolution: 1 Hz
Accuracy: ±(1.5 % of reading + 4 digits) on
Accuracy: ±(0.1 % of reading + 3 digits)
400 µA to 300 mA ranges
±(2.5 % of reading + 4 digits) on10 A range
Sensitivity: 10 Hz to 4MHz: >1.5 V rms;
4 MHz to 40 MHz: >2 V rms, <5 V rms
Peak hold accuracy: ±(3.0 % + 60 digits)
Min pulse width: >25 ns
Burden voltage: See DC Current
Duty cycle limits: >30 % and <70 %
Input protection: 0.315 A/1000 V fast blow
ceramic fuse 6.3×32 mm on µA/mA input
Overload protection: 1000 V dc or 750 V ac
10 A/1000 V fast blow ceramic fuse
rms
10x38 mm on 10 A input
10 A Input: 10 A for 4 minutes maximum
followed by a 12 minute cooling period
10
www.
.com