21.2

Float for SC250 with hygienic design

Two special tools are necessary to disassemble the float (2) from the flowmeter. They can be
supplied on demand by Tecfluid S.A.
These tools fit into the side guides of the float (2) and of the float end (8).
Once the tools have been fitted into their respective guides, turn the float end (8) as if it was
a nut. That way the float (2) will be separated from the float end (8), and the float can be
taken out from the top of the meter and the float end from the bottom.
Clean the inside of the metering tube and the float (2).
To remove adhered chemical dirt, clean the parts with suitable products or solvents and soft
brushes, never use metallic tools.
Follow the above steps in reverse to reassemble the equipment, making sure that the O-ring
(10) is correctly placed between the float (2) and the float end (8). Tighten the float end (8)
moderately onto the float (2).
If you are interested in acquiring the special tools in Tecfluid S.A., please indicate for which
nominal diameter they are required.