TESTO 0638.1345 Handling Instructions

TESTO 0638.1345 Handling Instructions

Pressure probe, plus minus 10 hpa/100 hpa, 2 bar abs

Advertisement

Quick Links

99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Observe measuring range (max. excessive pressure), too high pressure values can dest-
Please read the appropriate instruction manual before using the instrument!
In order to obtain reproducable measurements with differential pressure probes and to
maintain the accuracy ±3Pa with the 10 hPa probe, the following should be observed:
- The probe must be adjusted to the ambient temperature before the start of the
measurement!
- Zero calibration must be carried out before beginning a measurement, the position of
the pressure probe should not be changed during the measurements that follow!
- The pressure probe can be positioned as required.
- Avoid changes in temperature during the measurement!
- Observe + and - inputs!
0638.1445 (±10 hPa), 0638.1545 (±100 hPa)
Please read before using instrument!
roy the sensor!
The pressure probe works with a differential pressure
according to the principle of a strain gauge.
The pressure inlets in the differential pressure probe are
marked with + and -. The higher pressure should be
applied to the + inlet to achieve the best possible
accurary level. Higher pressure at the + inlet produces
a positive pressure indication, while higher pressure at
the - inlet produces a negative pressure indication.
Set the connected probe to its position of use before
beginning a measurement. We recommend the use of a
magnetic holding fixutre (see ordering data). Zero calibrati-
on should be carred out before the
measurement in order to guarantee reproducable measu-
rements. The position of the probe should not be altered
during the measurement that follows! Zero calibration is
not necessary for measurements with the absolute pres-
sure probe (0638.1645). The pressure probes are calibra-
ted in the factory, calibration is carried out when the pres-
sure probe is lying flat (normal position). We therefore
recommend that the probe remains in this position during
measurements.
Make sure that the pressure probe is not subject to vibra-
tions during the measurement (vibrations effect the results
of the measurement).
Handling instructions
Pressure probe
0638.1645 (2 bar abs)
Description
Connecting the probes
Application

Advertisement

loading

Summary of Contents for TESTO 0638.1345

  • Page 1 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Handling instructions Visit us at www.TestEquipmentDepot.com Pressure probe 0638.1445 (±10 hPa), 0638.1545 (±100 hPa) 0638.1645 (2 bar abs) Please read before using instrument! Observe measuring range (max. excessive pressure), too high pressure values can dest- roy the sensor! Please read the appropriate instruction manual before using the instrument! In order to obtain reproducable measurements with differential pressure probes and to...
  • Page 2: Technical Data

    Make sure that no liquid gets into the connec- tion nipple! Ordering data Pressure sensor ±100 Pa ......0638.1345 Pressure sensor ±10 hPa ......0638.1445 Pressure sensor ±100 hPa ......0638.1545 Pressure probe 2bar abs ......0638.1645 Magnetic holding fixture.......0554.0225...

This manual is also suitable for:

0638.14450638.16450638.1545