13.3 GUIDANCE AND MANUFACTURER'S DECLARATION—IMMUNITY
The T7 is intended for use in the electromagnetic environment specified below. The T7 user should ensure that it is used in
such an environment.
The T7 has been certified to the safety and EMC requirements of IEC 60601-1, 2nd and 3rd edition.
Immunity Test
ESD IEC
61000-4-2
EFT IEC
61000-4-4
Surge IEC
61000-4-5
Voltage Dips/
Dropout IEC
61000-4-11
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
All Equipment and Systems
50
IEC 60601 Test Level
±6 kV Contact
±8 kV Air
±2 kV Mains
±1 kV I/Os
1 kV Differential
±2 kV Common
> 95% Dip Cycle for .05 Cycle
60% Dip Cycle for 5 Cycles
30% Dip for 25 Cycles
95% Dip for 5 seconds
3 A/m
Compliance Level
±6 kV Contact
±8 kV Air
±2 kV Mains
I/Os N/A
±1 kV Differential
±2 kV Common
> 95% Dip Cycle for .05 Cycle
60% Dip Cycle for 5 Cycles
30% Dip for 25 Cycles
95% Dip for 5 seconds
3 A/m
Electromagnetic
Environment—Guidance
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the
r/h should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the T7 requires continued
operation during power
mains interruptions, it is
recommended that the
T7 be powered from an
uninterruptable power
supply or battery.
Power frequency magnetic
fields should be that
of a typical commercial or
hospital environment.