  • Page 1 Quick Start Guide IAA300 Differential Strain Gauge Amplifier Sensor Solutions Source Load · Torque · Pressure · Multi-Axis · Calibration · Instruments · Software www.futek.com...
  • Page 2: Getting Help

    Getting Help TECHNICAL SUPPORT For more IAA300 support, please visit: http://www.futek.com/iaa/support.aspx SP1177 FUTEK reserves the right to modify its design and specifications without notice. Please visit http://www.futek.com/salesterms for complete terms and conditions. 10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 www.futek.com...
  • Page 3: Table Of Contents

    Table of Contents Getting Help ..... . . 2 Power Side Connections ....6 Logic Board Configuration .
  • Page 4: Logic Board Configuration

    Logic Board Configuration Excitation Polarity Gain Bandwidth Shunt Selection Shunt Button Span Zero Note: Remove the magnetic cover to gain access to the logic board.
  • Page 5: Sensor Side Connections

    Sensor Side Connections IMPORTANT NOTE: Do not connect the device to the power supply when the power supply is already on. SENSOR SIDE PIN # WIRING CODE SHIELD + EXCITATION + SIGNAL – SIGNAL – EXCITATION For 6 wire sensors, connect +SENSE to +EXCITATION and –SENSE to –EXCITATION. Note: Sensor cable shield connections should be grounded on one end, either the sensor side or the IAA sensor input side, to avoid potential ground loops.
  • Page 6: Power Side Connections

    Power Side Connections IMPORTANT NOTE: Do not connect the device to the power supply when the power supply is already on. POWER SIDE PIN # WIRING CODE + Vin Shunt + Vout – Vout Shield Power is 12.5VDC to 26VDC. Note: Output is differential.
  • Page 7: Bandwidth Setup

    Bandwidth Setup IMPORTANT NOTE: Do not connect the device to the power supply when the power supply is already on. SW4 BANDWIDTH BANDWIDTH (Hz) ▲ ▼ ▼ ▼ ▼ ▲ ▼ ▼ 1,000 ▼ ▼ ▲ ▼ 10,000 ▼ ▼ ▼...
  • Page 8: Setup Steps

    With the sensor and IAA amplifier completely con- span output. nected apply the 12.5 to 26 VDC. FUTEK’s online calibration tool allows you to Note: Adjusting the Span will affect any system calibrations. retrieve a summary of your sensor’s calibration data: Adjusting the zero will not.
  • Page 9 5 to 26 VDC. Shunt Shunt Button Selection The online Shunt calculator on the FUTEK website can be used to calculate an estimated result from a shunt resistance, or to calculate a resistance needed for a certain sensor output value when shunted. http://www.futek.com/shuntcalc.aspx...
  • Page 10 10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 www.futek.com...