i-CAT FLX Installation Manual

i-CAT FLX Installation Manual

Cone beam 3d+ 2d panoramic dental imaging system
Hide thumbs Also See for FLX:
Table of Contents

Advertisement

TM
TM
C o n e B e a m 3 D +
2 D P a n o r a m i c D e n t a l
I m a g i n g S y s t e m
I n s t a l l at i o n M a n u a l

Advertisement

Table of Contents
loading

Summary of Contents for i-CAT FLX

  • Page 1 C o n e B e a m 3 D + 2 D P a n o r a m i c D e n t a l I m a g i n g S y s t e m I n s t a l l at i o n M a n u a l...
  • Page 2 FLX Installation Manual IMPORTANT! . . . X-RAY PROTECTION X-ray equipment may cause injury if used improperly. The instructions contained in this manual must be read and followed when installing this scanner. The scanner provides a high degree of protection from unnecessary X-radiation. However, no practical design can provide complete protection, nor prevent operators from exposing themselves or others to unnecessary radiation.
  • Page 3: Table Of Contents

    Alternate Network Configurations ............... 3-10 Scenario 1 - i-CAT FLX Installation ..............3-11 Scenario 2 - Replace Existing i-CAT 17-19 or KaVo 3D eXam with an i-CAT FLX (without DEXIS) ................3-12 Scenario 3 - New DEXIS - i-CAT FLX Installation ........3-13...
  • Page 4 FLX Installation Manual Scenario 4 - Upgrade Existing DEXIS (without i-CAT 17-19 or KaVo 3D eXam) with DEXIS - i-CAT FLX ..........3-15 Install and Configure Clinical Software ............3-17 Install SmartScan STUDIO Server Components ..........3-17 Install SmartScan STUDIO Client Components ..........3-21 Install TxSTUDIO Software ...................
  • Page 5 Table of Contents Scanner Shutdown ..................B-2 Log out ........................B-2 Power Off ........................ B-2 Appendix C - Setup of Alternative Network Configurations Remote Image Root Setup for Domain Environment ........C-1 Remote Network Share Using Windows Workgroup Environment ... C-6 032-0330-EN Rev K...
  • Page 6 FLX Installation Manual 032-0330-EN Rev K...
  • Page 7: Supplemental Components

    Supplemental Information Supplemental Components Patient E-stop Carbon Fiber Head Rest Part # 1304-0 Part # 27-0 Quantity 1 Quantity 1 Glide Head Restraint Band Part # 910-24 Part # 27-1 Quantity 4 Quantity 50 Tool Kit Velcro Head Restraint Part # 910-22 Part # 903-0 Quantity 1 Quantity 1...
  • Page 8: Pan Scan Components

    FLX Installation Manual Cord Retention Overhead Cord Bracket Bracket Part # 1.012.5468 Part # 1.012.6471 Quantity 1 Quantity 1 GeoCal Fixture QA Phantom Part # 14-1-0 Part # 13-00 Quantity 1 Quantity 1 Water Jar Phantom Chin Rest Slide, Straight...
  • Page 9: Chapter 1 - Scanner Assembly

    Chapter Scanner Assembly NOTE: Do NOT begin installation of any equipment or software unless the Preinstall Checklist (ISI-MKTG- CS-0007) has been completed for the site. Install Scanner There are two methods used for shipping the scanner, assembled and non-assembled. Start with one for the non-assembled scanner.
  • Page 10 FLX Installation Manual Remove both top covers (six mounting screws). 3/32 WARNING This scanner requires a two person lift. Failure to comply may cause bodily injury. Two adults are required to assemble this scanner. CAUTION CAUTION Use extreme care not to scratch overhead assembly when mounting scanner onto leg assemblies.
  • Page 11 Scanner Assembly When seated properly, overhead locks into place. To ensure safety, the overhead should still be supported by an assistant until mounted. From rear of scanner, secure overhead with mounting hardware, two SHCS1/4-20 x 5/8 long with 1/4” split washers, both sides. 3/16 The scanner must be checked for squareness prior to installing the mounting hardware at the...
  • Page 12 FLX Installation Manual From top of scanner, secure overhead with mounting hardware, two SHCS 1/4-20 x 5/8 long with 1/4” split washers, both sides. 3/16 CAUTION CAUTION Care must be taken when sliding the scanner into place. To prevent the scanner from scratching the floor surface, place the four glides under the scanner feet.
  • Page 13 Scanner Assembly Remove four long screws from red shipping plate. Keep shipping plate attached with remaining hardware. Do NOT discard mounting hardware. Long screws used for receptor assembly mounting and short screws for source assembly. WARNING The receptor assembly requires a two person lift. Failure to comply may cause bodily injury. Locate receptor assembly and remove packing material.
  • Page 14 FLX Installation Manual CAUTION CAUTION When mating connectors, do not force-fit connectors together. All connectors are color coded and keyed to prevent improper mating. Feed gray wire with red/white connector through brackets and connect to other red/white connector. 24. Feed second gray wire through brackets and connect the red/black wires with white connectors.
  • Page 15 Scanner Assembly Unlatch panel cover by pulling cover forward slightly, then swinging cover outward to remove. 28. Feed LEMO connector through hub and plug into panel. Rotate red dot on connector to the 3 o’clock position. Slide connector into receptor until it snaps in place. 29.
  • Page 16 FLX Installation Manual Carefully detach and move inner source cover a few inches away from assembly. 36. Remove ribbon cable from beam limiter assembly by pressing connector locking tab. Using very little force, slowly rotate gantry, counterclockwise. Stop the rotation when you feel that the gantry is at the limit switch position.
  • Page 17 Scanner Assembly With an assistant, align X-ray source assembly mounting pins. 40. Attach assembly using the four mounting screws previously removed from red shipping plate along with four split washers. 41. Draw the 3 cables out through the opening. Ensure 3/16 that the cables are not stretched or damaged.
  • Page 18 FLX Installation Manual Slide inner source cover into place, onto top mounting notches. 48. Mount outer cover onto mounting notches, over panel cover. 49. Attach covers with hex mounting screws at top and bottom (quantity 4 outer cover and 4 inner source cover).
  • Page 19: Install Retention Brackets

    Scanner Assembly Connect power cable to AC POWER IN connector on rear of overhead panel. 57. Ensure the power circuit breaker on the overhead rear panel is set to the 0 (off) position. To ensure that the device cabling does not interfere with the rotation of the gantry, use cable clips to dress cabling as shown.
  • Page 20 FLX Installation Manual 1. Determine proper bracket set configuration to use based on outlet receptacle and wall plug that will be used. See figure and use 3/16 configuration A or B. 2. Plug scanner line cord into wall receptacle.
  • Page 21: Install Scanner Controller, Touch Screen And Keyboard

    Scanner Assembly Install Scanner Controller, Touch Screen and Keyboard 1. Place scanner controller, touch screen and keyboard in position. 2. Connect power cables for scanner controller and for touch screen to each device and to power strip. For scanner controller and touch screen, use a power strip that is rated at a minimum of 2350 joules.
  • Page 22: Level Gantry

    FLX Installation Manual Level Gantry To achieve optimal performance, it is imperative that the gantry be properly leveled. 1. Scanner must be leveled from side-to-side and front-to-back. Level Side-to-Side Level Front-to-Back To level Overhead Gear, adjust the front gantry feet from the top, using a 1/4”...
  • Page 23 Scanner Assembly Store red shipping bracket in overhead. Replace shielding cover. Replace top cover (6 screws). 3/32 Install mounting bar on wall, centered behind the scanner. The height should be approximately 66 ½ to 68 inches [169 to 173 cm] from the floor to the center of the mounting bar.
  • Page 24: Check Version Of Smartscan Studio

    Protocol screen. If the protocols displayed are not correct for the field of view purchased, for example, V8 protocols display instead of V17 protocols, contact Technical Support to obtain an updated license file. Refer to the i-CAT FLX User Manual for a list of the protocols for each field of view type.
  • Page 25 Scanner Assembly The webpage URLs may need to be changed depending on the region of the world where the scanner is being installed. The default settings are for the United States. 1. Log in to the scanner controller and select Set Support URLs from the menu. 2.
  • Page 26 FLX Installation Manual 032-0330-EN Rev K 1-18...
  • Page 27: Chapter 2 - Calibration And Laser Adjustments

    Chapter Calibration and Laser Adjustments Calibrations are accessed from the Utilities menu in SmartScan STUDIO. See Service Menu Options chapter for instructions on logging in and accessing Utilities. NOTE: Ensure scanner is mounted to the wall before beginning calibrations. This should have been completed during the Level Gantry procedure.
  • Page 28: Panel Calibration

    FLX Installation Manual Panel Calibration It is recommended that Panel Cal be performed once a week. The Panel Cal is performed in both landscape and portrait (V17 systems only) positions for 4 x 4 and 2 x 2 resolutions. Several tests run as part of the panel calibration. A pie chart displays status.
  • Page 29 Calibration and Laser Adjustments ADJUSTMENTS Horizontal NOTCH Vertical Rotation NOTCH NOTCH NOTCH CAUTION CAUTION Ensure that the cover is positioned so that the side notches are toward the bottom of the panel. Remove beam limiter cover to gain access to the crosshair laser.
  • Page 30 FLX Installation Manual To move horizontal laser line up/down, turn adjustment screw (shown): • Clockwise - moves line up • Counter-clockwise - moves line down Vertical Line Adjustment: 5/32 4. Check position of vertical crosshair line with NOTCH panel cover notches.
  • Page 31: Patient Chair Alignment

    Calibration and Laser Adjustments Rotate Crosshair Laser Lines: 7. Check crosshair lines with panel cover notches to see if a rotation adjustment is required. To rotate laser crosshairs slightly loosen the two laser mounting screws (shown). 9. To help rotate the laser, insert an allen wrench (3/32) into the top hole of the laser.
  • Page 32: Run Chair Calibration

    FLX Installation Manual • If the Vertical Laser is out of position, adjust the chair foot from the top, using a 1/4 inch nut driver. Have a second person tilt the scanner enough to minimize the load on the foot while turning.
  • Page 33: Adjust Centerline Laser

    Calibration and Laser Adjustments Centering Patient Chair: 8. Select 90 and press 9. When prompted, press Scan button on control box. An audible alarm is sounded and the X-ray ON light is illuminated during radiation exposure. 10. Review the scout image. 11.
  • Page 34 FLX Installation Manual There are three laser adjustments: • Right-to Left • Angle • Laser Line Sharpness RIGHT-TO-LEFT ANGLE Right-to-Left Adjustment: 1. Loosen laser assembly set screw which allows the assembly to rotate. 2. Press the ALIGNMENT LIGHT button.The laser lights for approximately two minutes.
  • Page 35: Geometric Calibration

    Calibration and Laser Adjustments Angle Adjustment: 3/32 1. Loosen laser pointer set screw. 2. Press the ALIGNMENT LIGHT button. 3. While the laser is lit, manually rotate the laser pointer until it is aligned with the center line on the chair calibration fixture.
  • Page 36 FLX Installation Manual 5. If the Geometric Calibration Processing Failure message is displayed, the detector pivot is out range. The error message displays the allowed value (+/-0.050 degrees) and the actual value. To adjust detector pivot in landscape position: a.
  • Page 37: Shutter Calibration

    Calibration and Laser Adjustments a. On the receptor assembly, loosen the 5/32” set screw. b. The 1/4” adjustment screw is used to adjust the detector pivot. Turning the adjustment screw one full turn is an adjustment of 0.3. Typically, turn screw counter-clockwise to add to the displayed number and clockwise to subtract from the displayed number (facing the screw).
  • Page 38: Run Shutter Calibration

    FLX Installation Manual Run Shutter Calibration 1. From Utilities menu, select ShutterCal. 2. Ensure the field of view on the scanner is clear. 3. Press . The scanner initializes. 4. When prompted, press the Scan button on the operator control box. An audible alarm is sounded and the X-ray ON light is illuminated during radiation exposure.
  • Page 39: Chapter 3 - Load Clinical Software And Configure System

    Sites can have varied hardware configurations. At a minimum, there must be one server, in addition to the i-CAT FLX scanner controller, that can serve as the long-term data storage device and as a client workstation. Most sites will have multiple workstations, where one server stores long-term data and one or more clients are attached to the server.
  • Page 40 • PACS MWL • Legacy Practice Management Viewer Launcher • DEXIS i-CAT FLX Client • NOTE: The DEXIS i-CAT FLX Client supports the following operating systems: ® Windows 7 Professional, Ultimate, and Enterprise (64-bit and 32-bit) SP1 • ® Windows 8.1 (64-bit and 32-bit)
  • Page 41 Load Clinical Software and Configure System The following diagrams show the possible installation topologies for both an i-CAT FLX installation and a DEXIS installation. Refer to Appendix C for additional network setup options. i-CAT FLX Installation Shared Server - Client...
  • Page 42 FLX Installation Manual DEXIS - i-CAT FLX Installation (New or Existing) Shared Server - Client DEXIS Core*/DEXIS Eleven SmartScan STUDIO Scanner with Integration Services SmartScan STUDIO DEXIS i-CAT FLX Server Plugin DEXIS i-CAT FLX Scanner Controller Client with software pre-installed...
  • Page 43: Site It Network Setup

    Server Read/write access to ImageRoot “Committed” images do not Import Service folder appear in ImageRoot folder ImageRoot contains thumbnails, DEXIS i-CAT FLX SmartScan STUDIO Read/write access to DEXIS but DEXIS Core does not Server Plugin Server Core server Thumbnail does not appear in...
  • Page 44: Smartscan Studio Manager Status Indicators

    SmartScan STUDIO Manager displays three status indicators in the top, right-hand corner of the display. Move the mouse over the indicator to display more detail about the status condition. Scanner - indicates status of the connectivity between the workstation running SmartScan STUDIO Manager and the i-CAT FLX scanner controller. Database - •...
  • Page 45 Explorer. - User permissions to connect to Image Make sure Image Root folder is shared and Root folder incorrect or insufficient password protection removed. See i-CAT FLX Installation Manual, Appendix C for setup of alternative network configurations. 032-0330-EN Rev K...
  • Page 46: Fixed Ip Addresses

    SmartScan STUDIO Manager and Integration Services have to maintain continuous connection with the i-CAT FLX scanner controller. It is highly recommended that a fixed IP address, either static IP or DHCP Reservation, is assigned to the scanner controller to ensure a continuous connection. It is the responsibility of the site IT department to determine which method they want use to assign IP addresses.
  • Page 47: Risk Of Changes To Site Network

    Risk of Changes to Site Network The site IT department should be aware of the following: 1. Connection of the i-CAT FLX to the customer network/data coupling that includes other equipment could result in previously unidentified risks to patients, operators, or third parties.
  • Page 48: Alternate Network Configurations

    FLX Installation Manual 3. Subsequent changes to the network/data coupling could introduce new risks and require additional analysis. 4. Changes to the network/data coupling include: • changes in network/data coupling configuration. • connection of additional items to the network/data coupling.
  • Page 49: Scenario 1 - I-Cat Flx Installation

    If not already loaded, load on the workstation where the site will be viewing QA Tests. Load on workstations that will be using it to view patient images, as purchased by the site. NOTE: Vision is not supported on the i-CAT FLX, only TxSTUDIO and Dolphin viewers are supported.
  • Page 50: Scenario 2 - Replace Existing I-Cat 17-19 Or Kavo 3D Exam With An I-Cat Flx (Without Dexis)

    Scenario 2 - Replace Existing i-CAT 17-19 or KaVo 3D eXam with an i-CAT FLX (without DEXIS) This installation scenario is similar to an i-CAT FLX installation except that the existing ImageRoot must be moved to the site-provided server prior to installing the i-CAT FLX. Use the following sequence:...
  • Page 51: Scenario 3 - New Dexis - I-Cat Flx Installation

    If not already loaded, load on the workstation where the site will be viewing QA Tests. Load on workstations that will be using it to view patient images, as purchased by the site. NOTE: Vision is not supported on the i-CAT FLX, only TxSTUDIO and Dolphin viewers are supported.
  • Page 52 NOTE: For a Shared Server - Client installation scenario, you must also load the DEXIS i-CAT FLX Client on the server (located on Client Components media). For a Separate Server - Client installation scenario, load the DEXIS i-CAT FLX Client component on the server only if the site plans to use the server as a client.
  • Page 53: Scenario 4 - Upgrade Existing Dexis (Without I-Cat 17-19 Or Kavo 3D Exam) With Dexis - I-Cat Flx

    NOTE: DEXIS is optional software and is not available in all countries. Sites must be running DEXIS Eleven to be able to integrate with the i-CAT FLX and SmartScan STUDIO. For sites that are running earlier versions of DEXIS, follow the sequence below. For sites already running DEXIS Eleven, skip step 2.
  • Page 54 NOTE: For a Shared Server - Client installation scenario, you must also load the DEXIS i-CAT FLX Client on the server (located on Client Components media). For a Separate Server - Client installation scenario, load the DEXIS i-CAT FLX Client component on the server only if the site plans to use the server as a client.
  • Page 55: Install And Configure Clinical Software

    PACS Storage Service - (for sites using a PACS for image storage) • Data Utility (optional) • Patient Data Utility (optional) • Legacy Practice Management Interface - (for sites using a supported PM system) • DEXIS i-CAT FLX Server Plugin - (for sites using DEXIS) • 032-0330-EN Rev K 3-17...
  • Page 56 FLX Installation Manual NOTE: The Legacy Practice Management Interface and the DEXIS i-CAT FLX Server Plugin cannot be installed on a computer at the same time. b. Click Install. c. Click Yes on the confirmation dialog. The discovery service prepares for installation. If the required .NET framework and Visual C++ packages are not installed, the required files will...
  • Page 57 Load Clinical Software and Configure System 9. If the DEXIS i-CAT FLX Server Plugin was selected to be installed: a. On Welcome screen, click Next. b. On DEXIS configuration screen, use the default folder path. Click Next. c. On Confirm Installation screen, click Next. Installation will start and a progress bar is displayed.
  • Page 58 FLX Installation Manual a. Open Control Panel and select SmartScan STUDIO Integration Services. b. Check that Integration Services Connection shows name (computer name is default) and address of the Integration Services server. You can enter a unique name (will not change the computer name).
  • Page 59: Install Smartscan Studio Client Components

    Load Clinical Software and Configure System If DEXIS is installed, select next to Practice URL, then select the Practice URL from the dialog. Enter the Username and Password for the server, then select Test to ensure the connection is established. Click OK on the pop-up. g.
  • Page 60 Legacy Practice Management Viewer Launcher - (for sites using a supported PM • system) DEXIS i-CAT FLX Client - (for sites using DEXIS) • b. Click Install. c. Click Yes on the confirmation dialog. The discovery service prepares for installation. This may take a few minutes.
  • Page 61 Load Clinical Software and Configure System 8. If the DEXIS i-CAT FLX Client was selected to be installed: a. On DEXIS i-CAT FLX Configuration screen, click Next. b. On Ready to Install screen, click Install. Installation will start and a progress bar is displayed.
  • Page 62: Install Txstudio Software

    Manager Status Indicators (page 3-6) and troubleshoot. Click and close SmartScan STUDIO Manager. 11. For DEXIS sites, check settings on DEXIS i-CAT FLX Configuration: a. Open Control Panel and select DEXIS i-CAT FLX. b. Check Integration Services Connection options. Automatic By Name (default)
  • Page 63 Load Clinical Software and Configure System 3. Load TxSTUDIO on all workstations that will be used to view patient images: NOTE: TxSTUDIO installations are limited to the number of licenses held by the site (default of 6). a. At clinical workstation, insert media in drive. Access the TxSTUDIO installer from the AutoPlay dialog box.
  • Page 64: Configure Smartscan Studio Pacs Modules (Optional)

    FLX Installation Manual Configure SmartScan STUDIO PACS Modules (Optional) For sites that use a PACS (Picture Archiving Communication System) for storing images, or use a RIS (Radiology Information System) to maintain patient data, configure the optional SmartScan STUDIO PACS software on the clinical workstations.
  • Page 65 Load Clinical Software and Configure System 5. In the Storage Commitment section, select the Perform Storage Commitment checkbox if you want to enable storage commitment of sent DICOM images. When selected, two radio buttons become active: Select Use Existing Connection to use the same connection for sending a Storage •...
  • Page 66 FLX Installation Manual 3. Click Options. a. Under Integration Service Connection, select one of the following: Automatic by Name (Default) - This is • the preferred option. Connects to SmartScan STUDIO Integration Services server location by computer name. This prevents connection problems caused by changing IP addresses.
  • Page 67 Load Clinical Software and Configure System 2. Click Config (PACS Server configuration settings): a. Enter Station name (user defined for identifying a particular PACS server setup). Enter IP Address, AE Title, and Port number for the remote PACS server (obtain from PACS administrator), from where studies are to be queried and retrieved.
  • Page 68: Configure Legacy Practice Management Interface (Optional)

    Configure Legacy Practice Management Interface (Optional) For sites that use a Practice Management system, configure the optional PM Interface software on the server. Refer to the i-CAT FLX User Manual for information about status messages that can be displayed. NOTE: •...
  • Page 69: Installation Check

    6. Click Save & Restart on SmartScan STUDIO Integration Services. Installation Check Installation Check for i-CAT FLX Installation 1. At clinical workstation, enter data for a test patient and schedule exam. 2. At scanner controller, capture a scan for test patient.
  • Page 70: Installation Check For A Dexis I-Cat Flx Installation

    4. In DEXIS, locate Test Patient exam in Extra-oral display. 5. Double-click the i-CAT box to start TxSTUDIO and load exam. 6. View scan in TxSTUDIO and ensure correct path to the TxSTUDIO Cases folder displays in File Manager.
  • Page 71: Chapter 4 - Qa Tests

    Chapter QA Tests Run QA Tests Run the following QA Tests. See Service Menu Options chapter for instructions on logging in and accessing Utilities. • QA Line Pair Test • QA Material Test • QA Air Water Test • QA Pan Test CAUTION CAUTION If the optional deadman handswitch is installed on the scanner, press and hold handswitch before...
  • Page 72: Run Qa Line Pair Test

    FLX Installation Manual 4. Using the Alignment Lasers, adjust the platform height so that the horizontal laser is positioned at the center of the QA phantom. Make sure the phantom is centered left to right and front to back. Use the lasers to confirm.
  • Page 73: Qa Line Pair Evaluation

    QA Tests QA Line Pair Evaluation 1. At a clinical workstation, start SmartScan STUDIO Manager and select Exam List. 2. On Exam List, locate QA 3D_1 Line Pair exam with most current date. Double-click CT entry to load study in TxSTUDIO. NOTE: If a pop-up message displays stating “Image data is too small”, click OK and continue.
  • Page 74: Distance Measurement Test

    FLX Installation Manual 9. Verify that definition is present within line pairs 10, 11, and 12. Compare image quality to the image shown. Distance Measurement Test To ensure measurement accuracy, this procedure checks Distance measurements. 1. Select to activate Distance tool.
  • Page 75: Qa Material Evaluation

    QA Tests 3. Select , then press . The scanner initializes. 4. When prompted, press the Scan button on the operator control box. An audible alarm is sounded and the X-ray ON light is illuminated during radiation exposure. 5. Review the scout image. The phantom must be centered and level. Adjust the phantom platform as needed to achieve the proper height.
  • Page 76 FLX Installation Manual 5. Zoom upper left image. To zoom, move the mouse cursor in the center of the image, hold down the Control key and press the left mouse button. Move the cursor up or down to zoom in or out as needed.
  • Page 77: Qa Air Water Test

    QA Tests 11. Record the Mean HU value of each material. See table below. Recorded values should fall within the lower and upper limits. Mean Scan Value Lower Upper Mean Material (Hounsfield Units) Limit Limit Air (Black) (lower left) -1000 -980 -990 Acrylic (Light Gray) (lower right)
  • Page 78: Run Qa Air Water Test

    FLX Installation Manual 2. Fill phantom half full with distilled water and carefully place on platform. 3. Using the Alignment Lasers, center the water bath with the horizontal laser across the center of the water depth. Run QA Air Water Test 1.
  • Page 79 QA Tests NOTE: If a pop-up message displays stating “Tru-Pan failed to process”, click OK and continue. 3. When exam is loaded, select Section tab at top of display. 4. In the upper right corner view, click where the vertical and horizontal cursor lines cross and drag to the center of the water height and width, as shown.
  • Page 80 FLX Installation Manual 9. Record the Water HU Mean value in the table below. 10. In the upper right image, slide the vertical and horizontal cursor lines above the water level. 11. Record the Air HU Mean value in the table below. Recorded values should fall within the ranges.
  • Page 81 QA Tests Uniformity Evaluation The Uniformity test is used to check that image density measurements are consistent in all areas within the field of view. 13. In the upper right corner view, click where the vertical and horizontal cursor lines cross and drag to the center of the water height and width.
  • Page 82: Qa Pan Test

    FLX Installation Manual QA PAN Test Install PAN Phantom 1. Prepare the bite tip by inserting the narrow edges of the bite tip down into the bite tip holder uprights. Then turn the bite tip a ¼ turn to lock into place.
  • Page 83: Qa Pan Test Evaluation

    QA Tests 7. When prompted, press the Scan button on the operator control box. An audible alarm is sounded and the X-ray ON light is illuminated during radiation exposure. The scanner acquires data and a status indicator shows acquisition and image creation progress. When image processing is complete, image is displayed.
  • Page 84 FLX Installation Manual 032-0330-EN Rev K 4-14...
  • Page 85: Chapter 5 - Complete Installation Forms

    Chapter Complete Installation Forms Be sure to complete and return the following forms, if applicable: FDA 2579 - Report of Assembly of a Diagnostic X-ray System I-6800-56 - i-CAT FLX Installation Checklist 032-0330-EN Rev K...
  • Page 86 FLX Installation Manual 032-0330-EN Rev K...
  • Page 87: Wall Mount With Cable Exposed

    Appendix Wall Mounting Operator Control The operator control box is configured for desktop usage but can be modified for wall mounting access. Desktop Wall Mounted Wall Mounted (cable exposed) (cable inside wall) Wall Mount with Cable Exposed Disconnect operator control box cable from rear of overhead panel.
  • Page 88 FLX Installation Manual Lift side panel (tab end) upward and remove from operator control box (both sides). CAUTION CAUTION A ribbon cable is connected from the operator control box face panel to the base. Damage to the ribbon cable may occur if the face panel is separated too far from the base.
  • Page 89 Wall Mounting Operator Control Box Rotate face panel 180 8. Connect ribbon cable as shown. Ensure that connector is fully seated. Mount face panel with the four mounting screws. There are four outer mounting rings on the operator control box base. These rings can be pre-drilled or punctured with a screw during mounting.
  • Page 90: Wall Mount With Cable Inside Wall

    FLX Installation Manual Wall Mount with Cable Inside Wall Disconnect operator control box cable from rear of overhead panel. Remove cable connector cover plate from rear of operator control box. Pull cable out from wire channel. 4. Press the plug locking latch.
  • Page 91 Wall Mounting Operator Control Box Protect wires and plug by wrapping them in electrical tape prior to feeding through wall. The plug requires at least a 1” [25mm] diameter hole. The wall hole location should be approximately 1” [25mm] below the top of the operator control box location.
  • Page 92 FLX Installation Manual At top of operator control box, slide clip away from tab so that tab can be pressed down to loosen side panel (both sides). Lift side panel (tab end) upward and remove from operator control box (both sides).
  • Page 93: Appendix B - Service Menu Options

    Exam List Access utility to add, update, delete and view status of exams. Access user account management, network information, file Configurator maintenance, and export dose logbook and activity logs. Refer to i-CAT FLX Technical Guide. Technical Support Access Technical Support website.
  • Page 94: Run Utilities

    QA Line Pair Refer to QA Tests chapter. QA Material QA Air Water QA Pan Reprocess Exam Refer to i-CAT FLX User Manual. Favorites Manager Refer to i-CAT FLX User Manual. Roll-off Refer to i-CAT FLX Service Manual. Scanner Shutdown Log out 1.
  • Page 95: Appendix C - Setup Of Alternative Network Configurations

    Appendix Setup of Alternative Network Configurations The following procedures describe setup needed to support the following alternative network configurations. You may need to work with the site’s IT representative to complete a procedure. • Remote Image Root Setup for Domain Environment •...
  • Page 96 FLX Installation Manual e. Click Object Types button, then click to check the Computers check box. Click Enter the domain and machine name of the server running SmartScan STUDIO Integration Services. NOTE: The machine name must end in a dollar sign. Example: gendex\hpaltehughes$ g.
  • Page 97 Setup of Alternative Network Configurations 2. Configure Windows Service Log On The remaining steps are performed on the SmartScan STUDIO Integration server. Update the SmartScan STUDIO Integration Services to use the Network Service account for logon so that the network share can be accessed. 1.
  • Page 98 FLX Installation Manual 5. Enter Network Service in object names field, and click OK. 6. Grant full access to the folder. Select Allow check box for Full Control and Read options. Click OK. 4. Local Folder Security Configuration Grant the Network Service account full control permissions of the following folders: •...
  • Page 99 Setup of Alternative Network Configurations 3. Use the following command to verify that the reservations exist: netsh http show urlacl Typical response should include the following entries: Reserved URL :https//+:8585/VisionRoot/ User: NT AUTHORITY\NETWORK SERVICE Listen: Yes Delegate: No SDDL:D:(A;;GX;;;NS) Reserved URL :http//+:8484/VisionRoot/ User: NT AUTHORITY\NETWORK SERVICE Listen: Yes...
  • Page 100: Remote Network Share Using Windows Workgroup Environment

    FLX Installation Manual Remote Network Share Using Windows Workgroup Environment Use this procedure to enable access to a network shared Image Root folder on a remote network. 1. Configure Remote Network Folder for Storing Image Root Data The remote Image Root computer is a Windows-based file server that provides a network folder for storing image root data.
  • Page 101 Setup of Alternative Network Configurations e. Enter SSSService in object name field and click OK. Grant full access to SSSService account. Select Allow check box for Full Control, Change, and Read options. Click OK to close any open windows. The network share should now be accessible to other computers on the network.
  • Page 102 FLX Installation Manual b. On the Member Of tab, click Add. c. Enter Administrators in object name field, then click OK to close any open windows. The user is now a member of the Administrators group. 3. Update the SmartScan STUDIO Integration...
  • Page 103 Setup of Alternative Network Configurations 4. Update the image root to the image root server using the SmartScan STUDIO control panel. a. Start Control Panel and double-click SmartScan STUDIO Integration Services control panel. b. On Update Service tab, browse to the remote Image Root directory.
  • Page 104 FLX Installation Manual 032-0330-EN Rev K C-10...
  • Page 106 Manufactured by: Imaging Sciences International LLC 1910 North Penn Road Hatfield, PA 19440 USA Tel: 1-215-997-5666 Fax: 1-215-997-5665 032-0330-EN Rev K 2016 August 1...

Table of Contents