janitza UMG 96S 24V Operating Instructions Manual
janitza UMG 96S 24V Operating Instructions Manual

janitza UMG 96S 24V Operating Instructions Manual

Universal measuring device
Table of Contents

Advertisement

Quick Links

Janitza electronics GmbH
Janitza electronics GmbH
Janitza electronics GmbH
Vor dem Polstück 6
Vor dem Polstück 6
Vor dem Polstück 6
D-35633 Lahnau
D-35633 Lahnau
D-35633 Lahnau
Support Tel. 0049 6441 9642-22
Support Tel. 0049 6441 9642-22
Support Tel. 0049 6441 9642-22
Fax 0049 6441 9642-30
Fax 0049 6441 9642-30
Fax 0049 6441 9642-30
e-mail: info@janitza.com
e-mail: info@janitza.com
e-mail: info@janitza.com
Internet: http://www.janitza.com
Internet: http://www.janitza.com
Internet: http://www.janitza.com
Universal Measuring Device
Universal Measuring Device
Universal Measuring Device
UMG 96S 24V
UMG 96S 24V
UMG 96S 24V
Operating instructions
Operating instructions
Operating instructions
Brief instructions see last page
Brief instructions see last page
Brief instructions see last page
Max value,
Max value,
Max value,
HT/Imported supply
HT/Imported supply
HT/Imported supply
Min value,
Min value,
Min value,
Supply
Supply
Supply
LT/Supply
LT/Supply
LT/Supply
Mean value
Mean value
Mean value
Programming mode
Programming mode
Programming mode
Sum measurement
Sum measurement
Sum measurement
Phase-to-phase
Phase-to-phase
Phase-to-phase
Password
Password
Password
Voltage transformer
Voltage transformer
Voltage transformer
Current transformer
Current transformer
Current transformer
Output 1
Output 1
Output 1
Output 2
Output 2
Output 2
Key 1
Key 1
Key 1
Key 2
Key 2
Key 2

Advertisement

Table of Contents
loading

Summary of Contents for janitza UMG 96S 24V

  • Page 1 Universal Measuring Device Universal Measuring Device Universal Measuring Device UMG 96S 24V UMG 96S 24V UMG 96S 24V Operating instructions Operating instructions Operating instructions Brief instructions see last page Brief instructions see last page Brief instructions see last page Max value,...
  • Page 2: Table Of Contents

    Contents Meaning of the symbols used Parameters and Measured Values 34 Notes on Use Parameter display at the UMG96S Inspection on Receipt Measured value display at the UMG96S 34 Programming parameters Scope of supply Mean values Maintenance Instructions Current averaging time (Add 057) Repair and calibration Power averaging time (Add 058) Front film...
  • Page 3 Tables Parameter list Measured value list Measured value displays, overview Display Ranges and Accuracy Technical Specifications Ambient conditions Declaration of conformity Safety requirements EMC requirements Test voltages (type testing) Inputs and outputs Auxiliary voltage Measurement Serial interfaces Connectable conductors Dimensioned drawings Connection examples Quick Reference Instructions Changing the current transformer...
  • Page 4: Meaning Of The Symbols Used

    Meaning of the symbols used The symbols used in this instruction manual have the following meaning: Warning, dangerous electric voltage. This symbol is intended to warn you of possible hazards that can occur during installation, starting up and use. All rights reserved. No part of this manual may be reproduced or duplicated without the written approval of the author.
  • Page 5: Notes On Use

    Notes on Use Inspection on Receipt This device may only be deployed and used by Fault free and safe use of this device requires qualified personnel in accordance with the appropriate transport, proper storage, erection safety instructions and regulations. When using and assembly as well as careful operation and the device, any additional legal and safety regu- maintenance.
  • Page 6: Maintenance Instructions

    07:00 to 12:00 hrs age temperature of +45°C. The typical life ex- pectancy of the battery is 8 to 10 years. Janitza electronics GmbH For safety reasons, the battery can only be re- Vor dem Polstück 1 placed in the manufacturer’s factory!
  • Page 7: Product Description

    Product Description Intended use Important! The UMG96S is intended for fixed, permanent The neutral conductor N must always and weather protected use in switchboards and be connected. for the measurement of voltage, current, power, etc. in low voltage switchgear. The measure- ment is designed for 3-phase systems with Important! neutral conductor (TN and TT systems).
  • Page 8: Functional Principle

    Functional principle The three-phase electronic measuring system records and digitises the effective values of al- ternating voltages and alternating currents in 50Hz/60Hz systems. ../5A or ../1A current transformers can be op- tionally connected to the current measurement inputs. In systems with voltages up to 150VAC to earth, currents up to 5A can also be directly 18-70V DC connected to the UMG96S.
  • Page 9 Page 9...
  • Page 10: Device Versions

    Device versions The UMG96S is available in various design versions. In these the terminals 11, 12 and 13 at the UMG96S can be assigned a func- tion specified by the customer. Measuring- I / O Design version 1 circuit voltage Design version 1 contains the following func- tional groups: Input/Output...
  • Page 11 Design version 2 Design version 2 contains the following functional groups: RS232 RS232 (MODBUS RTU) Input/Output Measuring- I / O Pulse output 1 circuit voltage (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Current measurement UMG96S Master board...
  • Page 12 Design version 3 Design version 3 contains the following functional groups: RS232 RS232 (MODBUS RTU) Input/Output Measuring- circuit voltage Pulse output 1 (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Clock with battery Data memory Current measurement UMG96S...
  • Page 13 Design version 4 Design version 3 contains the following functional groups: RS232 RS232 (MODBUS RTU) Input/Output Measuring- circuit voltage Pulse output 1 (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Analog output 1 Analog output 2 Current measurement UMG96S...
  • Page 14 Design version 5 Design version 5 contains the following functional groups: RS232 RS232 (MODBUS RTU) Input/Output Measuring- Pulse output 1 circuit voltage (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Analog output 1 Analog output 2 Current measurement Clock with battery...
  • Page 15 Design version 6 Design version 6 contains the following functional groups: RS232 (MODBUS RTU) RS232 Input/Output Measuring- Pulse output 1 circuit voltage (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Digital input 1 Digital input 2 Current measurement UMG96S...
  • Page 16 Design version 7 Design version 7 contains the following RS232 functional groups: RS232 (MODBUS RTU) Measuring-circuit and operating Profibus DP voltage Input/Output Pulse output 1 (Wp=active power demand) Pulse output 2 (Wq=reactive power demand) Digital output 1 Digital output 2 Current measurement Digital input 1 Profibus...
  • Page 17 Page 17...
  • Page 18: Installation Instructions

    Installation Instructions Installed position The UMG96S is intended for permanent instal- - The wiring cables for the operating voltage lation in low and medium voltage switchgear. It must be suitable for rated voltages up to can be installed in any position. 300VAC to earth.
  • Page 19: Current Measurement

    Current measurement Summation current measurement The current is optionally measured via a ../5A If the current is measured via two current trans- or ../1A current transformer. If, in addition to the formers, the total transformation ratio of the UMG96S, the current has to be measured with current transformers must be programmed in a ampere meter, this must be connected in se- the UMG96S.
  • Page 20: Serial Interfaces

    Serial interfaces Inputs and outputs The various design versions of the UMG96S In the various design versions (options) of the have up to two serial interfaces. UMG96S, different functions can be assigned The serial interfaces are not isolated from each to the outputs.
  • Page 21 230V AC External 230V AC Auxiliary voltage External Auxiliary 24V DC 24V DC voltage +24V= Digital input 1 Digital input 2 Analog output UMG96S Digital inputs Digital output 1.5k Fig.: Connection example for the digital inputs. UMG96S Connection example: UMG96S with one ana- log output and one digital output.
  • Page 22: Connection Options

    Connection options UMG96S UMG96S Voltage Current Voltage Current measurement measurement measurement measurement cf. rating plate 0.005 .. 5A cf. rating plate 0.005 .. 5A 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 L1 L2 L3 N 1k 1l 2k 2l 3k 3l L1 L2 L3 N...
  • Page 23 UMG96S UMG96S Voltage Current Voltage Current measurement measurement measurement measurement cf. rating plate 0.005 .. 5A cf. rating plate 0.005 .. 5A 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 L1 L2 L3 N 1k 1l 2k 2l 3k 3l L1 L2 L3 N...
  • Page 24: Putting Into Service

    Putting into Service Applying the auxiliary voltage The UMG96S should be put into service as fol- The size of the auxiliary voltage is given on the lows: rating plate. Auxiliary voltages which do not cor- respond to those given on the rating plate can Installing the device result in malfunctions and destruction of the The UMG96S is intended for installation in low-...
  • Page 25: Applying The Measuring-Circuit Voltage

    Applying the measuring-circuit voltage Programming the current and voltage The wiring cables for the measuring-circuit transformers voltages to the UMG96S must be suitable for A current transformer of 5/5A is set in the fac- voltages up to 300V to earth and 520V phase- tory.
  • Page 26: Applying The Measuring Current

    Applying the measuring current Checking the phase assignment The UMG96S is designed for the connection of The assignment of the phase conductor to the ../1A and ../5A current transformers. current transformer is correct if you short-circuit Only alternating currents, not direct currents, a current transformer on the secondary side can be measured via the current measurement and the current displayed by the UMG96S in...
  • Page 27: Checking The Current Direction

    Checking the current direction Checking the individual power outputs Short-circuit two current transformers on the If a current transformer is assigned to the secondary side. The active power displayed in wrong phase conductor, the corresponding the remaining phase of the UMG96S must now: power output will also be incorrectly measured be positive (+) when active power is imported and displayed.
  • Page 28: What To Do In Case Of Errors

    What to do in case of errors Possible error Cause Remedy Back-up fuse has triggered. Insert fuse. Display dark. Device defective. Send device to the manufacturer for repair. Measured value The measured value display Add the required measured value display cannot be has been deleted from the display to the measured value selection.
  • Page 29 Possible error Cause Remedy Phase shift ind/cap. Current path is assigned to Check connection and correct if the wrong voltage path. necessary. Programming data is The device has been exposed Improve external protective measures lost. to electromagnetic such as shielding, filtering, earthing and interferences, which are larger spatial separation.
  • Page 30: Error Messages

    Error Messages The UMG96S displays three different error Example: Error number 911 messages in the display: The UMG96S displays the error number 911. - Warnings, - fatal errors and - Overranges. In the event of warnings and fatal errors the error message is represented by the symbol “EEE”...
  • Page 31: Warnings

    Warnings Overrange Warnings are less serious errors and can be Overranges are displayed as long as they exist acknowledged using Key 1 or Key 2. The re- and cannot be acknowledged. An overrange cording and display of measured values contin- exists if at least one of the three voltage or cur- ues.
  • Page 32: Operation And Display

    Operation and Display Programming mode The UMG96S is operated using the Keys 1 and 2. Measured values and programming data are shown on a liquid crystal display. A differentia- tion is made between display mode and ϕ ϕ ϕ ϕ ϕ MkWh MkVArh Programming mode...
  • Page 33: Key Functions

    Key functions Display mode Password Programming mode simultaneously simultaneously Programming Measured menu values Programming Measured menu values Programming Measured Measured menu values values long short Programming Confirm selection menu short Digit +1 long Digit -1 short Value *10 (decimal point to the right) long Value /10...
  • Page 34: Parameters And Measured Values

    Parameters and Measured Parameter display at the UMG96S Values In this example the value „0001“ is shown as parameters required operate being the contents of address „000“ in the dis- UMG96S, e.g. the current transformer data, play of the UMG96S. Here the UMG96S has and all measured values are stored in a list.
  • Page 35: Programming Parameters

    Programming parameters Simultaneously press both keys for around 1 second. If a user password has been programmed the password query appears with “000”. The first digit of the user password flashes and can be changed using Key 2. If you press Key 2 the next digit is selected and flashes.
  • Page 36: Mean Values

    Mean values Min and max values Mean values for the measured current and All the measured values are measured and cal- power output values are formed over an adjust- culated once a second. Min and max values are able period of time. The mean values are de- determined for most of the measured values.
  • Page 37: System Frequency (Add 063)

    System frequency (Add 063) Active power demand In the UMG96S the system frequency is deter- The UMG96S has four work meters. Three ac- mined from the measuring-circuit voltage of tive power demand counters and one reactive phase L1. The scanning frequency for the cur- power demand counters.
  • Page 38: Current Transformer (Add 600)

    Current transformer (Add 600) Current transformers with a secondary current Programming of either 1A or 5A can be optionally connected In programming mode, use Key 2 to page to the to the UMG96S. current transformer setting. Use Key 1 to con- A current transformer of 5A/5A is programmed firm the selection.
  • Page 39: Voltage Transformer (Add 602)

    Voltage transformer (Add 602) Programming The voltage phase conductor to phase con- In programming mode, use Key 2 to page to the ductor (L-L) is given as the secondary and pri- voltage transformer setting. Use Key 1 to con- mary voltage in the display of the UMG96S. firm the selection.
  • Page 40: Harmonics (Add 221)

    Harmonics (Add 221) Total harmonic distortion THD (Add 269) Harmonics (harmonic components) are the in- The total harmonic distortion calculated in the teger multiple of a fundamental component. UMG96S for current and voltage gives the ratio The UMG96S measures the fundamental com- of the effective value of the distortion magni- ponent of the voltage within the range 45 to tude to the effective effective value of the peri-...
  • Page 41: Measured Value Paging

    Measured value paging All the measured values are calculated once a Display change profiles (Add 061) second and can be called up in the measured Adjustment range: 0 .. 3 value displays. Two methods are available for 0 - display change profile No. 0, pre-assigned. calling up the measured value displays: 1 - display change profile No.
  • Page 42 In the overview of the measured value displays, „A01“ corresponds measured values voltages L-N. Display change profile No. 1 Display change profile No. 2 (cf. also Page 86 to 89) (cf. also Page 86 to 89) C D E C D E Page 42...
  • Page 43 Display change profile No. 3 (Add 605) The customer-specific display change profile No.3 can only be configured using the PC software GridVis and not directly at the UMG96S. This requires a connection between UMG96S and PC via a serial interface (RS232 or RS485).
  • Page 44: Measured Value Displays

    Measured value displays After a system recovery, the UMG96S displays Displays profile (Add 060) the first measured value table from the current Adjustment range: 0 .. 3 displays profile. In order to keep selection of the 0 - displays profile No. 0, measured values to be displayed manageable fixed pre-assigned default.
  • Page 45 Measured value displays profile (Add 604) The customer-specific measured value dis- plays profile No. 3 can only be configured using the PC software GridVis and not directly at the UMG96S. This requires a connection between UMG96S and PC via a serial interface (RS232). Format of the measured value displays profile: STRING Byte 1 = row 1,...
  • Page 46: User Password (Add 011)

    User password (Add 011) Delete work (Add 009) A user password can be programmed to make The UMG96S has four work meters. Three ac- it difficult for the programming data to be acci- tive power demand counters and one reactive dentally changed.
  • Page 47: Phase Sequence (Add 277)

    Phase sequence (Add 277) LCD contrast (Add 012) The phase sequence of the voltages and the The preferred viewing direction for the LCD dis- frequency of phase L1 are shown in a display. play is from „below“. The LCD contrast of the The phase sequence indicates the phase se- LCD display can be adjusted by the user.
  • Page 48: Time Recording

    Time recording Serial number (Add 911) The UMG96S records the operating hours of The serial number displayed by the UMG96S the UMG96S and the total running time of each has 6 digits and is part of the serial number dis- comparator.
  • Page 49: Software Release (Add 913)

    Software release (Add 913) Hardware configuration (Add 914) The software for the UMG96S is continuously The options available in the UMG96S can be improved and enhanced. The software status in queried via the address 914. One bit is set for the device is denoted with a 3-digit number, the each available option.
  • Page 50: Serial Interfaces

    Serial interfaces The various design versions of the UMG96S RS232/Modbus have up to three serial interfaces. Profibus DP RS232 RJ11 socket The serial interfaces are not isolated from each other. The profibus interface can be run simul- taneously with the RS232 interface. Modem mode (Add 070) An analog modem can be connected to the UMG96S via the RS232 interface.
  • Page 51: Modbus Rtu

    MODBUS RTU Example: Reading out the voltage L1-N The voltage L1-N is stored in the measured The data from the parameter and measured value list under the address 200. The voltage value list can be accessed via the MODBUS L1-N is stored in INT format. RTU protocol.
  • Page 52: Rs232 Interface

    RS232 interface The achievable distance between two RS232 devices depends on the cable used and the baud rate. As a guideline measurement, for a transmission rate of 9600 baud the distance should not exceed 15m to 30m. The permissible ohmic load must be larger than 3 kOhm and the capacitive load caused by the transmission cable must be smaller than 2500 With the PC cable for the RS232 interface (2m)
  • Page 53 Connection examples GridVis UMG96S PC cable (2m) Product number 08.01.501 RS232/DSUB-9 RS232/RJ11 Fig. Connect the UMG96S with a PC via a PC cable. GridVis Modem cable (2m) Product number 08.01.503 UMG96S Modem Modem RxD TxD RxD TxD RS232/RJ11 Fig. Connect UMG96S with a PC via modem. PC cable Modem cable DSUB -9...
  • Page 54: Profibus Dp

    Profibus DP The UMG96S has a 9-pin SubD socket in the Device master file rear panel. A RS485 interface with which the The device master file for the UMG96S is called profibus DP protocol is operated is wired to the „U96S0781.GSD“.
  • Page 55 Cable length The maximum permissible cable length depends on the type of cable and the transmission rate level. The length of the cable is measured between the bus driver of the first device and the bus driver of the last device. We recommend you only use shielded cables which conform to the standard EN50170-2 cable type A.
  • Page 56: Profibus Profiles

    Profibus profiles Very many measured values are available in the Profile formats UMG96S for further processing. In order to keep The measured values in the 14 profiles are in the amount of data to be transferred via the integer format and can be called up in floating decimal point format.
  • Page 57 „Profibus profiles“ lists, integer format Profibus profile No. 1 Profibus profile No. 2 Profibus profile No. 3 Profibus profile No. 4 Value Bytes Value Bytes Value Bytes Value Bytes Uln L1 eeprom_timer I_with_L1 Uln L2 Comp_timer_1 I_with_L2 Uln L3 Comp_timer_2 I_with_L2 UL1-L2 Comp_timer_3...
  • Page 58 Profibus profile No. 8 Profibus profile No. 9 Profibus profile No. 10 Profibus profile No. 11 Value Bytes Value Bytes Value Bytes Value Bytes thd_i_L1 thd_u_L1 tdh_i_L1 dft_i_9_L1 thd_i_L2 thd_u_L2 thd_i_L2 dft_i_9_L2 thd_i_L3 thd_u_L3 thd_i_L3 dft_i_9_L3 dft_i_1_L1 dft_u_1_L1 thd_u_L1 dft_u_9_L1 dft_i_1_L2 dft_u_1_L2 thd_u_L2...
  • Page 59 Profibus profile No. 12 Profibus profile No. 13 Profibus profile No. 14 Value Bytes Value Bytes Value Bytes P_total_max thd_i_L1_max thd_u_L1_max P_total_max_with 2 thd_i_L2_max thd_u_L2_max I_total_max thd_i_L3_max thd_u_L3_max I_total_max_with 2 dft_i_1_L1_max dft_u_1_L1_max 2 phi_total_max 2 dft_i_1_L2_max dft_u_1_L2_max 2 S_total_max dft_i_1_L3_max dft_u_1_L3_max 2 Q_total_max dft_i_3_L1_ma...
  • Page 60: Inputs And Outputs

    Inputs and outputs The terminals 12 and 13 at the UMG96S can be optionally assigned the following functions: 0 = Pulse output, 1 = Digital output, Voltage I / O 2 = Analog output (option), measurement 3 = Digital input (option), 4 = Profibus remote output (option), 5 = HT/LT switchover via a Digital input (option),...
  • Page 61 Status displays The status of the switching inputs and outputs is denoted in the UMG96S’s display by circular symbols. Status at terminal 12 Status at terminal 13 Statuses at the digital input: A maximum voltage of 2V is applied. A voltage larger than 20V is applied. The status of the digital input is not defined for voltages within the range from 2V to 20V.
  • Page 62: Pulse Output

    Pulse output Each digital output can be used as a pulse The pulses collected within a second are out- put with the programmed pulse length and a output. Pulse output 1 can only output the maximum frequency of 10Hz. active power demand Wp and pulse output 2 If the measured work (power demand) exceeds can only output the reactive power demand the set pulse value, so that the maximum fre-...
  • Page 63: Pulse Value

    Minimum pulse length (Add 010) Pulse value The minimum pulse length can be adjusted in The pulse value denotes how much work 10ms steps within a range from 50ms up to (power demand - Wh or kvarh) a pulse equals. 1000ms.
  • Page 64 Example: Programming the pulse output The UMG96S is to measure measure the ac- 1.) Select pulse output tive power demand in a sub-distribution and for- Assign terminal 12 the pulse output function. warded to a data collector via the pulse output. Input/Output 1 Add 002=0 The sub-distribution supplies consumers which together require a maximum 400kW active...
  • Page 65 3.) Determining the pulse value The maximum work (power demand), which can be imported in an hour is: Work = active power demand * time Work = 400kW *1h Work = 400kWh If a pulse value of 400kWh per pulse is set, at full load the UMG96S supplies one pulse.
  • Page 66: Digital Output

    Digital output Two digital outputs can be assigned to termi- nals 12 and 13 of the UMG96S. The value 001 must be written to the parameter address 002 for digital output 1 and 001 to parameter ad- dress 003 for digital output 2. The result of limit value monitoring (388, 392) is then output at the corresponding digital out- put.
  • Page 67 Example: Current monitoring in the N If the current in the N is greater than 100A for 230V AC External 60 seconds, digital output 1 should cut in for at Operating least 2 minutes. voltage 24V DC The following programming steps have to be +24V= undertaken: 1st Comparator...
  • Page 68: Limit Value Monitoring

    Limit value monitoring Two comparator groups, each with 3 compara- Important! tors, are available for monitoring limit values. Only the first three digits of a parameter The results of comparators A, B and C can be can be set at the UMG96S. AND or OR gated and the result can optionally All the digits of a parameter can be ad- be inverted.
  • Page 69 Measured value (Add 015) Operator (Add 017) The measured address contains the address of Two operators are available for comparing the the measured value to be monitored. The fol- measured value and limit value. lowing values can be assigned to the measured Operator = 0 corresponds to greater than value: or equal (>=)
  • Page 70: Analog Output

    Analog output The UMG96S with the additional printed circuit Scale start value (Add 050, 054) board 1 has 2 analog outputs. Each analog out- The scale start value equals the measured put can output a current of 4-20mA. An exter- value at which the minimum current of 4mA is nal 24VDC power supply unit is required for its to flow.
  • Page 71 230V External Operating 24V DC voltage +24V= Analog output 4 - 20mA Analog output 4 - 20mA UMG96S Analog outputs Fig. Connection of the analog outputs to a programmable controller. 230V External Operating 24V DC voltage +24V= Analog output 4 - 20mA Analog output Analog recorder...
  • Page 72: Digital Input

    Digital input The UMG96S with the additional printed circuit Status of the digital inputs If a digital input is assigned the function „3“, the board 2 has 2 digital inputs. An external 24VDC status of the digital inputs can be queried via power supply unit is required for operation of the addresses 420 and 421 .
  • Page 73 HT/LT switchover (Add 071) If a digital input is assigned the function „5“, this digital input can be used to switch between the active power demand meter (kWh register) HT and the active power demand meter (kWh reg- ister) LT. If no voltage is applied at the digital input, a 0 is written in address 071.
  • Page 74: Memory

    Memory Data recording (056) In the UMG96S, one EEPROM memory is al- The measured values and results available to ways available for the configuration data and choose from for storage are summarised in 4 min and max values. In addition, a data pre-defined memory profiles.
  • Page 75 Memory profile 1 The mean power output values are summa- rised in memory profile 1: Mean value P in L1 Mean value P in L2 Mean value P in L3 Mean value Q total Mean value S total The data in memory profile 1 is always saved after the averaging time P has expired.
  • Page 76: Tables

    Tables Parameter list The parameter list contains all the settings required for correct operation of the UMG96S, e.g. current transformer and device address. The values in the parameter list can be written and read. Data and time in address 700 are an exception and can only be written. The date and time in seconds since 1.1.1970 can be read in address 410 of the measured value list.
  • Page 77: Parameter List

    Parameter list Add Name Adjustm. range Type Units Default UMG96S Device address 1 .. 255 CHAR Baud rate (RS232 and RS485) 0 .. 2 CHAR kbps Input/Output 1, type 0 .. 5 CHAR Input/Output 2, type 0 .. 5 CHAR Pulse value, digital output 1 0 ..
  • Page 78 Parameter list Part 2 Name Adjust.. range Type Units Default Comparator 2C, measured value 0 .. 999 INT 6) Comparator 2C, minimum on-time 1 .. 899 Sec. Comparator 2C, operator 0, 1 CHAR 4) Gate the results of the comp. (0,1,2) 0, 1 CHAR 1) Invert result of logic oper.
  • Page 79 Parameter list Part 3 Name Adjust. range Type Units Default Current transformer, primary3) 1 .. 10000 Current transformer, secondary3) 1 .. 5 Voltage transformer, primary3) 100 .. 60000 4001) Voltage transformer, secondary3) 100 .. 400 4001) Measured value displays profile, current STRING1 Display change profile, current STRING1...
  • Page 80: Measured Value List

    Measured value list Name Type Units Resolution Voltage L1-N Voltage L2-N Voltage L3-N Voltage L1-L2 Voltage L2-L3 Voltage L3-L1 Current in L1 Current in L2 Current in L3 Active power demand L1 Active power demand L2 Active power demand L3 Reactive power demand L1 Reactive power demand L2 Reactive power demand L3...
  • Page 81 Measured value list Part 2 Name Type Units Resolution 9th harmonic I L1 11th harmonic I L1 13th harmonic I L1 15th harmonic I L1 1st harmonic I L2 3rd harmonic I L2 5th harmonic I L2 7th harmonic I L2 9th harmonic I L2 11th harmonic I L2 13th harmonic I L2...
  • Page 82 Measured value list Part 3 Name Type Units Resolution Mean value S, total Max value I, total Max value, P mean value, total Max value I mean value, total Max value, P total Max value, Q total Max value, S total Max value, CosPhi total 0,01 Min value, U L1-N...
  • Page 83 Measured value list Part 4 Name Type Units Resolution Max value, 11th harmonic U L2-N Max value, 13th harmonic U L2-N Max value, 15th harmonic U L2-N Max value, 1st harmonic U L3-N Max value, 3rd harmonic U L3-N Max value, 5th harmonic U L3-N Max value, 7th harmonic U L3-N Max value, 0th harmonic U L3-N Max value, 11th harmonic U L3-N...
  • Page 84 Measured value list Part 5 Name Type Units Resolution Comparator result 1A CHAR Comparator result 1B CHAR Comparator result 1C CHAR Total result of logic operation, comparator group 1 CHAR Comparator result 2A CHAR Comparator result 2B CHAR Comparator result 2C CHAR Total result of logic operation, comparator group 2 CHAR...
  • Page 85 Measured value list Part 6 Name Type Units Resolution Data memory data record FDATA Data memory data record FDATA Data memory data record FDATA Data memory data record FDATA Data memory data record FDATA Data memory data record FDATA 500-505 Decrement when reading out the data memory pointer 702 Data memory data record FDATA...
  • Page 86: Measured Value Displays, Overview

    Measured value displays, overview Measured values Max values Min values L1-N voltage L1-N voltage L1-N voltage L2-N voltage L2-N voltage L2-N voltage L3-N voltage L3-N voltage L3-N voltage Measured values Max values Min values L1-L2 voltage L1-L2 voltage L1-L2 voltage L2-L3 voltage L2-L3 voltage L2-L3 voltage...
  • Page 87 Mean value Max value (ind) Measured value Total reactive po- Total reactive po- Total reactive po- wer demand wer demand wer demand Measured value Measured value Measured value Distortion factor THD Distortion factor THD Distortion factor THD U L1 U L2 U L3 Measured value Measured value...
  • Page 88 Meas. value (ind) Total Reactive power demand Comparator 6 Operating hours Comparator 1 Total running time meter 1 Total running time Measured value Measured value Measured value 1st harmonic 15th harmonic 3rd harmonic U L1 U L1 U L1 Measured value Measured value Measured value 3rd harmonic...
  • Page 89 Max value Max value Max value 1st harmonic 3rd harmonic 15th harmonic U L1 U L1 U L1 Max value Max value Max value 1st harmonic 3rd harmonic 15th harmonic U L2 U L2 U L2 Max value Max value Max value 3rd harmonic 15th harmonic...
  • Page 90: Display Ranges And Accuracy

    Display Ranges and Accuracy Measured variables Display range Measured range 1) Measuring accuracy3) Voltage L-N 300V standard version 0 .. 34kV 50 .. 300V +-0.5% vMb 150V special version 0 .. 17kV 25 .. 150V +-0.5% vMb Voltage L-L 300V standard version 0 .. 60kV 87 ..
  • Page 91: Technical Specifications

    Technical Specifications Weight : 250g Calorific value : 2.2MJ (610Wh) Ambient conditions Overvoltage category : CATIII Degree of soiling Protection class : II = without protective conductor Operating temperature range : -10°C. +55°C Storage temperature range : -20°C. +70°C Relative air humidity : 15% to 95% without condensation Degree of protection Front...
  • Page 92: Test Voltages (Type Testing)

    Technical specifications Part 2 Test voltages (type testing) 300V standard version Measuring-circuit voltage to serial interfaces, inputs and outputs :3700V AC Current measurement inputs to serial interfaces, inputs and outputs :2300V AC The serial interfaces are not isolated from each other! 150V special version Measuring-circuit voltage to serial interfaces, inputs and outputs...
  • Page 93: Auxiliary Voltage

    Technical specifications Part 3 Auxiliary voltage Voltage range :18V .. 60V DC, 18V .. 30V AC 50/60Hz Power input :max. 1.4W, max. 2.6VA Back-up fuse : 2A. 10A (medium time lag) Measurement Measuring rate : 1 measurement/sec. Impulse voltage withstand level : 4kV Signal frequency : 45Hz.
  • Page 94: Dimensioned Drawings

    Dimensioned drawings +0.8 +0.8 Auxiliary section dimensions: 92 x 92 Rear Option Option Switchboard 1 2 3 4 1 2 3 4 Side view max. 6 1 2 3 4 1 2 3 4 1 2 3 4 1 2 3 4 1 2 3 4 Fixing bracket 1 2 3 4...
  • Page 95: Connection Examples

    Connection examples UMG96S Current measurement Auxiliary Measuring-circuit voltage 0.005 .. 5A voltage input RJ11 11 12 13 14 15 1 2 3 4 5 6 7 8 9 10 L1 L2 L3 N 1k 1l 2k 2l 3k 3l 2 .. 10A ../5(1)A ../5(1)A ../5(1)A...
  • Page 96: Quick Reference Instructions

    Quick Reference Instructions Changing the current transformer Programming mode Switching to programming mode If you are in display mode and press Keys 1 and 2 for around one second, you switch to programming mode. The symbols for programming mode PRG and for the current transformer CT appear.

Table of Contents