Use Of Pid Lamp Cleaning Kit A-31063 - ION FirstCheck+ 2000 User Manual

Hide thumbs Also See for FirstCheck+ 2000:
Table of Contents

Advertisement

FirstCheck+ 1000, 2000, 3000 & 5000
Maintenance
3. Push a lamp-removing sleeve over the end of the lamp.
This allows the user to grip the glass.
4. Extract the lamp from its socket by pulling it in a straight
away from the PID cell. Force is only required when
pulling the lamp past the O ring. Once past the O ring
the lamp will move freely.
5. Inspection of the lamp may reveal a layer of
contamination on the detection window that presents
itself as a blue hue. To check for contamination, hold
the lamp in front of a light source and look across the
window surface.
WARNING!
NEVER REFIT A DAMAGED LAMP!

USE of PID lamp cleaning kit A-31063

1. The vial of cleaning compound contains Aluminium Oxide as a very fine power (CAS Number 1344-28-
1). A full material safety data sheet MSDS is available on request from Ion science Ltd. The key issues
are listed below:
Hazard identification:
May cause irritation of respiratory tract and eyes
Handling:
Do not breathe vapour/dust. Avoid contact with skin, eyes and clothing
Wear suitable protective clothing
Follow industrial hygiene practices: Wash face and hands thoroughly with soap and water after use and
before eating, drinking, smoking or applying cosmetics.
The Compound has a TVL(TWA) of 10 mg/m
Storage:
Keep container closed to prevent water absorption and contamination.
2. Open the vial of Aluminium Oxide polishing compound.
With a clean cotton bud collect a small amount of compound.
Ion Science Ltd
3
Page 37 of 45
Ion Science Ltd
Passed the O-
ring
Move freely in a
straight line